iBOT PMD with software version 01.05.24. Personal Mobility… recallZ-1965-2025 · Mobius Mobility LLC · initiated Apr 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mobius Mobility LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue that could potentially lead to the device tipping over from Balance Mode.

FDA root cause category: Software design

What the company is doing

Consignees were initially contacted via telephone on 4/24/25 to notify them of this software issue. An URGENT MEDICAL DEVICE CORRECTION NOTICE dated 5/5/25 was emailed to consignees to notify them of the recall and instructing them that the firm is developing a software correction. Once Mobius Mobility has obtained a correction they will reach out to consignees to schedule a time to update affected devices. Consignees with any questions can email service @mobiusmobility.com or call 833-346-4268x2.

Recall details

Recall number
Z-1965-2025
Status
Open, Classified
Initiated by the company
Apr 24, 2025
Posted by the FDA
Jun 12, 2025
Recalling firm
Mobius Mobility LLC
Quantity
442 units
Product code
IMK
FDA submissions
K243442
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, IN, NC, NJ, OH, PR, RI, TN, TX & VA.

Product description

iBOT PMD with software version 01.05.24. Personal Mobility Device.

Code information (models, lots, serial numbers)

Software Version 01.05.24; UDI (01) 00857584008010 (21) 0001108, (01) 00857584008010 (21) 0001110, (01) 00857584008010 (21) 0001121, (01) 00857584008010 (21) 0001123, (01) 00857584008010 (21) 0001128, (01) 00857584008010 (21) 0001136, (01) 00857584008010 (21) 0001137, (01) 00857584008010 (21) 0001422, (01) 00857584008010 (21) 0001438, (01) 00857584008010 (21) 0001496, (01) 00857584008010 (21) 0001505, (01) 00857584008010 (21) 0001508, (01) 00857584008010 (21) 0001516, (01) 00857584008010 (21) 0001517, (01) 00857584008010 (21) 0001519; Serial No. 0001108, 0001422, 0001438, 0001496, 0001505, 0001508, 0001516, 0001517, 0001519, 0001110, 0001121, 0001123, 0001128, 0001136, 0001137.