LOWER EXTREMITY, Model: NCEX20I recallZ-1916-2025 · American Contract Systems, Inc. · initiated Apr 21, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
FDA root cause category: Finished device change control
What the company is doing
On 4/21/2025, correction notices were emailed to the customer who was asked to do the following: 1) Discontinue use of the Cast Padding component. All other components within the kit can be used. 2) Add warning labels to affected kits in Table 2 to ensure the component is discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected Procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. - Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. - Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. You are also responsible for the reconciliation of these customers. Please do not have your customers respond directly to ACS. Complete and return the response form via email to [email protected] Report any quality problems or adverse events associated with the products listed in this notification to [email protected]
Recall details
- Recall number
- Z-1916-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2025
- Posted by the FDA
- Jun 6, 2025
- Recalling firm
- American Contract Systems, Inc.
- Quantity
- 850 Kits
- Product code
- OJH
- Distribution
- US Nationwide distribution in the state of CO.
Product description
LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain Specialist Blend Cast Padding, Item: QTX30-322NS
Code information (models, lots, serial numbers)
Model/UDI-DI/Lot(Expiration): NCEX20I/00191072222398/9146304(2025-12-03), 9207685(2026-01-08); NCEX20J/00191072234582/9222096(2026-01-30), 9222096(2026-01-30); NCTK31K/00191072222480/9086225(2025-11-01), 9086225(2025-11-01); NCTK31L/00191072234872/9219596(2026-01-15), 9219596(2026-01-15)
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS NCEX20IAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS NCTK31KAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS NCEX20JAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS NCTK31LAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗