Allon 2001. Part Number: 200-00263 recallZ-1966-2025 · Belmont Instrument LLC · initiated May 13, 2025

Class II Open, Classified All recalls posted in June 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Belmont Instrument LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.

FDA root cause category: Device Design

What the company is doing

Belmont Medical Technologies notified consignees via letter on about 05/13/2025. Consignees were instructed to ensure users of the devices are made aware of the correction, identify affected units on hand and check their serial numbers, ensure all devices have the Temp Sensor Alarms always set to ON to ensure that the user is alerted to errors and conditions requiring actions and attach the provided Step by Step Guide (P/N DDT-063-027) to each device. Consignees were also instructed to complete and return the response form and notify customers if the units were further distributed.

Recall details

Recall number
Z-1966-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 13, 2025
Posted by the FDA
Jun 12, 2025
Recalling firm
Belmont Instrument LLC
Quantity
229 units
Product code
DWJ
FDA submissions
K024128
Distribution
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.

Product description

Allon 2001. Part Number: 200-00263

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

Part Number: 200-00263. UDI-DI: 007290107581103, Serial Numbers: 99111143000002, 99111143000014, 99111627300034, 9993901600001, 9993901600002, 9993901600004, 9993901600006, 9993901600007, 9993901600009, 9995356300005, 9995356300006, 9995356300007, 9995356300008, 9996673800002, 99111143000001, 99111143000003, 99111143000004, 99111143000007, 99111143000008, 99111143000009, 99111143000011, 99111143000012, 99111143000013, 99111143000015, 99111143000018, 99111143000019, 99111288700016, 99111288700021, 99111288700035, 99111288700036, 99111627300011, 99111627300017, 99111627300022, 99111627300029, 99111627300040, 99111627300041, 99111627300045, 99111627300046, 99111627300054, 99111627300057, 99111627300064, 99111627300067, 99113168300004, 99113168300005, 99113168300008, 99113168300010, 99113168300012, 99128623400008, 99129614700002, 99129614700005, 99131046400001, 99131046400003, 99131046400004, 99131046400005, 99131046400006, 99131046400009, 99131046400011, 99131046500007, 99132732400009, 99132732400018, 99132732400019, 99134232100011, 99132121000010, 99113168300003, 9983349800008, 9991451200002, 9991451200004, 9992175000001, 9995356300010, 99111627300001, 99111627300008, 99113168300013, 99132121000014, 99132732400005, 99134232100001, 99134232100002, 99134232100003, 99134232100009, 99134232100022, 99134232100035, 99111627300042, 99111627300056, 99111627300058, 99111627300059, 99113168300014, 99111627300020, 99111627300049, 99113168300006, 99128623400003, 99132121000013, 99134232100018, 9991451200001, 99111143000006, 99111627300005, 99111627300035, 99113168300001, 99113168300002, 99113168300007, 99111627300053, 99111288700014, 99111288700017, 99111288700018, 99111288700019, 99111288700020, 99111627300016, 99111627300032, 99111143000017, 99111143000021, 99111143000022, 99111288700001, 99111288700002, 99129614700009, 9962076700011, 99132121000012, 99132121000015, 99132121000005, 99132121000007, 99111627300009, 99134232100017, 99131046400008, 9995356100003, 99113168300011, 99111627300025, 99132121000009, 99111627300061, 99111627300038, 9974083900001, 99111627300030, 9991451200006, 99111627300018, 9976534100006, 9945767700003, 9983349700008, 9991451200003, 9991451200009, 99111288700005, 99111627300031, 99128623400004, 99128623400005, 99128623400007, 99129614700001, 99129614700003, 99129614700004, 99129614700007, 99129614700008, 99131046400002, 99131046500001, 99131046500002, 99132121000001, 99128623400006, 99111143000016, 9969414000006, 99134232100013, 99134232100021, 99111627300036, 99132121000017, 99132121000018, 99132121000019, 99132121000020, 99132732400002, 99132732400006, 99132732400013, 99132732400014, 99134232100005, 99134232100010, 99132121000016, 99111627300013, 99111627300033, 99111627300043, 99111627300044, 99111627300051, 99111627300063, 99129614700006, 99111627300007, 99111627300024, 99111627300037, 99111627300047, 99111627300014, 99111627300052, 99113168300009, 99128623400001, 99128623400002, 99128623400009, 99128623400010, 99134232100008, 99111143000020, 99131046400010, 99132121000011, 99132732400007, 99132732400011, 99132732400012, 99132732400016, 99134232100034, 99132732400008, 99132732400015, 99132732400017, 99132732400020, 99134232100014, 9959060500014, 99111627300026, 99134232100004, ", 99134232100006", 9991451200005, 9995356200008, 99111143000010, 99111288700004, 99131046500003, 99131046500005, 9973536800007, 99131046500004, 99131046500008, 99131046500009, 99132121000004, 99132121000006, 99132121000008, 99134232100026, 99132732400001, 99132732400003, 99132732400004, 99134232100007, 99134232100019, 9940577300005, 99111143000005, 99111627300048, 99111627300062, 99132121000002, 9979841500003, 99132121000003, 99131046500010