Nellcor recallZ-2151-2025 · Covidien · initiated Jun 23, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

FDA root cause category: Labeling design

What the company is doing

On 6/23/2025 correction notices were distributed to customers who where informed: Per Operator s Manual (http://manuals.medtronic.com): 1) Device alarm volume needs to be set to a level appropriate to be heard throughout the care environment at all times of day or night. Do not silence or decrease the volume of the audible alarm if patient safety could be compromised. 2) If the sensor comes off the patient, or no perfusion is detected, the device will notify 'sensor off' (low priority alarm) every 16 seconds and display a steady yellow visual indicator. If a low priority alarm immediately follows a higher priority alarm, a user may not recognize the higher priority clinical condition. 3) Monitors should be set to Standard Mode when in use in a hospital or hospital-type environment by trained medical personnel and set to Homecare Mode when a lay person will be using the monitor outside of a hospital or other professional care setting. With the monitor not in Homecare Mode, alarms could be silenced/muted and alarm limits altered In Clinical Facility Environments be sure the PM100N model is set to Standard Mode for general use and only set to Sleep Study Mode when conducting a sleep study. Device should be set to Homecare Mode prior to use. Firm's Technical Service at 800-255-6774 (option 1, then 2) can provide a four-digit password required to set a monitor to Homecare Mode. Provide the Home Use Guide to home caregivers. Clinicians should communicate the recall information to home use patients and caregivers. Pass on/post this notice for all those who need to be aware within your organization and to any organization where the product has been transferred/distributed. Complete and return the Customer Confirmation Form to [email protected] Firm will provide additional information when finalized. Questions - Contact firm's Customer Service at 800-962-9888, Option 2.

Recall details

Recall number
Z-2151-2025
Status
Open, Classified
Initiated by the company
Jun 23, 2025
Posted by the FDA
Jul 21, 2025
Recalling firm
Covidien
Quantity
264,790
Product code
DQA
FDA submissions
K142865, K120773
Distribution
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.

Product description

Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

REF/UI-DI/Bundle Code/IFU Part Number: PM100N/10884521196728/NELL-HOMECARE, BNPULSOKSYMETR, PM100N, PM100NKIT, PM100NKITCHDE, PM100NKITCHFR, PM100NKITDA, PM100NKITDE-2, PM100NKITES, PM100NKITFI, PM100NKITFR, PM100NKIT-FR, PM100NKITIT, PM100NKIT-NL, PM100NKITNO, PM100NKITPT, PM100NKITSV, PM100NKITUK, PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC/PT00156589 - AR - Arabic, PT00156609 - AR - Arabic, PT00156588 - BG - Bulgarian, PT00156610 - BG - Bulgarian, PT00156586 - CS - Czech, PT00156612 - CS - Czech, PT00156585 - DA - Danish, PT00156613 - DA - Danish, PT00156578 - DE - German, PT00156625 - DE - German, PT00156577 - EL - Greek, PT00156626 - EL - Greek, PT00156509 - EN - English, PT00156328 - EN - English, PT00156563 - ES - Spanish, PT00156639 - ES - Spanish, PT00156583 - ET - Estonian, PT00156615 - ET - Estonian, PT00156582 - FI - Finnish, PT00156616 - FI - Finnish, PT00156581 - FR - French, PT00156617 - FR - French, PT00156587 - HR - Croatian , PT00156611 - HR - Croatian , PT00156576 - HU - Hungarian, PT00156627 - HU - Hungarian, PT00156575 - IT - Italian, PT00156628 - IT - Italian, PT00156580 - KA - Georgian, PT00156618 - KA - Georgian, PT00156591 - KO - Korean, PT00156642 - KO - Korean, PT00156573 - LT - Lithuanian, PT00156630 - LT - Lithuanian, PT00156574 - LV - Latvian, PT00156629 - LV - Latvian, PT00156584 - NL - Dutch, PT00156614 - NL - Dutch, PT00156572 - NO - Norwegian, PT00156631 - NO - Norwegian, PT00156571 - PL - Polish, PT00156632 - PL - Polish, PT00156570 - PT-BR - Brazilian Portuguese, PT00156633 - PT-BR - Brazilian Portuguese, PT00156569 - RO - Romanian, PT00156634 - RO - Romanian, PT00156567 - RU - Russian, PT00156635 - RU - Russian, PT00156565 - SK - Slovak, PT00156637 - SK - Slovak, PT00156564 - SL - Slovenian, PT00156638 - SL - Slovenian, PT00156566 - SR - Serbian, PT00156636 - SR - Serbian, PT00156562 - SV Swedish, PT00156640 -SV - Swedish, PT00156561 - TR - Turkish, PT00156641 - TR - Turkish, PT00156590 - ZH - SimplifIed Chinese, PT00156643 - ZH - SimplifIed Chinese; 10005941/10884521163454/PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC, BEDSIDESP2KIT2, NELLSPO2, PM100N-HC, NELL-HOMECARE, ROEBEDSIDEKIT, BEDSIDESP2PT2, BEDSIDESP2NO2, BEDSIDESP2PL2, BEDSIDESP2NL2, BEDSIDESP2KITNM, BEDSIDESP2K2UK, BEDSIDESP2K2IT, BEDSIDESP2K2FR, BEDSIDESP2K2-2, BEDSIDEKITISR/PT00156645 - AR - Arabic, PT00156647 - BG - Bulgarian, PT00156649 - CS - Czech, PT00156650 - DA - Danish, PT00156656 - DE - German, PT00156657 - EL - Greek, PT00156324 - EN - English, PT00156673 - ES - Spanish, PT00156652 - ET - Estonian, PT00156653 - FI - Finnish, PT00156654 - FR - French, PT00156648 - HR - Croatian, PT00156658 - HU - Hungarian, PT00156659 - IT - Italian, PT00156655 - KA - Georgian, PT00156677 - KO - Korean, PT00156661 - LT - Lithuanian, PT00156660 - LV - Latvian, PT00156651 - NL - Dutch, PT00156662 - NO - Norwegian, PT00156664 - PL - Polish, PT00156666 PT-BR - Brazilian Portuguese, PT00156667 - RO - Romanian, PT00156668 - RU - Russian, PT00156671 - SK - Slovak, PT00156672 - SL - Slovenian, PT00156669 - SR - Serbian, PT00156674 - SV - Swedish, PT00156675 - TR - Turkish, PT00156676 - ZH - Simplified Chinese, NELL-SPO2-AD3+ - Canadian, NELL-SPO2-AD3 - Canadian; 10005941/A8845211634501; DLPM100N/10884521527607; DL10005941/10884521173293; DSPM100N/DSPM100N; 10005941-SG/10884521171534/BNINHLLCODE, BNINNEL100A, BNINNEL100AS, BNINNEL100AVJ, BNINNEL100AYS, BNINNELDYS, BNHKLIBRANEOSET, BNHKLIBRASET; DS10005941/DS10005941; 10005941J/10005941JP; 10005941JP/884521512019; 10005941JPN/884521188689. All serial numbers