Medical device recalls: July 2025
The FDA posted 174 medical device recall records in July 2025, from 73 recall events by 66 companies. 21 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- MEDLINE INDUSTRIES, LP - Northfield · 2 events, 5 records
- Ortho-Clinical Diagnostics, Inc. · 2 events, 5 records
- Baxter Healthcare Corporation · 2 events, 4 records, 4 Class I
- Encore Medical, LP · 2 events, 4 records
- Medtronic, Inc. · 2 events, 4 records
- Microbiologics Inc · 2 events, 3 records
- Philips Medical Systems Nederland B.V. · 2 events, 2 records
- Philips Ultrasound, Inc · 1 event, 21 records
FDA root cause
- Process control · 42
- Under Investigation by firm · 41
- Device Design · 33
- Software design · 14
- Packaging change control · 14
- Radiation Control for Health and Safety Act · 6
- Process change control · 6
- Other · 4
Class I recalls (21)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2158-2025 · Draeger, Inc. · posted Jul 31, 2025 · Open, Classified
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
- Z-2159-2025 · Draeger, Inc. · posted Jul 31, 2025 · Open, Classified
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
- Z-2160-2025 · Draeger, Inc. · posted Jul 31, 2025 · Open, Classified
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
- Z-2161-2025 · Draeger, Inc. · posted Jul 31, 2025 · Open, Classified
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
- Z-2167-2025 · Hamilton Medical AG · posted Jul 30, 2025 · Open, Classified
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could…
- Automated Impella Controller (AIC), used in left heart support blood pump, including the following:Class IZ-2164-2025 · Abiomed, Inc. · posted Jul 26, 2025 · Open, Classified
A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
- Z-2095-2025 · Baxter Healthcare Corporation · posted Jul 17, 2025 · Open, Classified
Certain Spectrum infusion pumps may have an incorrect version of software.
- Z-2096-2025 · Baxter Healthcare Corporation · posted Jul 17, 2025 · Open, Classified
Certain Spectrum infusion pumps may have an incorrect version of software.
- Z-2102-2025 · Intersurgical Inc · posted Jul 11, 2025 · Open, Classified
Potential for faulty devices as a result of depleted batteries.
- Z-2047-2025 · Given Imaging Ltd. · posted Jul 10, 2025 · Open, Classified
It has been determined that misapplied adhesive on the Bravo CF capsule delivery device may lead to its malfunction. Specifically, the misapplied adhesive may prevent the capsule from attaching to the patient's esophagus or detaching from the delivery device.
- Z-2080-2025 · Edwards Lifesciences, LLC · posted Jul 10, 2025 · Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
- Z-2081-2025 · Edwards Lifesciences, LLC · posted Jul 10, 2025 · Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
- Brand Name: FEM-FLEX IIClass IZ-2082-2025 · Edwards Lifesciences, LLC · posted Jul 10, 2025 · Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
- Brand Name: FEM-FLEX IIClass IZ-2083-2025 · Edwards Lifesciences, LLC · posted Jul 10, 2025 · Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
- Brand Name: FEM-FLEX IIClass IZ-2084-2025 · Edwards Lifesciences, LLC · posted Jul 10, 2025 · Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
- Brand Name: FEM-FLEXIIClass IZ-2085-2025 · Edwards Lifesciences, LLC · posted Jul 10, 2025 · Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
- Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts:Class IZ-2031-2025 · Baxter Healthcare Corporation · posted Jul 9, 2025 · Open, Classified
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight…
- Z-2032-2025 · Baxter Healthcare Corporation · posted Jul 9, 2025 · Open, Classified
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight…
- Z-2035-2025 · Philips Respironics, Inc. · posted Jul 2, 2025 · Open, Classified
This device does not indicate for use in patients with respiratory failure.
- Z-2036-2025 · Philips Respironics, Inc. · posted Jul 2, 2025 · Open, Classified
This device does not indicate for use in patients with respiratory failure.
- Z-2037-2025 · Philips Respironics, Inc. · posted Jul 2, 2025 · Open, Classified
This device does not indicate for use in patients with respiratory failure.
Class II recalls (148)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2150-2025 · Apic Yamada America · posted Jul 31, 2025 · Open, Classified
Apic Yamada has recently discovered that certain products may be out of compliance with FDA electronic product regulations in that they did not bear a required label.
- Brand Name: OsteotecClass IIZ-2222-2025 · Osteotec Limited · posted Jul 31, 2025 · Open, Classified
There is the potential that the silicone implant may contain foreign material
- Z-2223-2025 · Remel, Inc · posted Jul 31, 2025 · Open, Classified
The products may contain surface and subsurface contamination of Listeria monocytogenes.
- Z-2224-2025 · Remel, Inc · posted Jul 31, 2025 · Open, Classified
The products may contain surface and subsurface contamination of Listeria monocytogenes.
- Z-2225-2025 · Remel, Inc · posted Jul 31, 2025 · Open, Classified
The products may contain surface and subsurface contamination of Listeria monocytogenes.
- Z-2226-2025 · Remel, Inc · posted Jul 31, 2025 · Open, Classified
The products may contain surface and subsurface contamination of Listeria monocytogenes.
- Z-2230-2025 · Siemens Healthcare Diagnostics Inc · posted Jul 31, 2025 · Open, Classified
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and…
- Z-2199-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jul 30, 2025 · Open, Classified
SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of…
- Achieva 3.0T ↗Class IIZ-2200-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Achieva XR ↗Class IIZ-2201-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Evolution Upgrade 1.5TClass IIZ-2202-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Evolution Upgrade 3.0TClass IIZ-2203-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia 1.5TClass IIZ-2204-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia 1.5T SClass IIZ-2205-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia 3.0TClass IIZ-2206-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia 3.0T CXClass IIZ-2207-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia Ambition SClass IIZ-2208-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia Ambition XClass IIZ-2209-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia Elition SClass IIZ-2210-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Ingenia Elition XClass IIZ-2211-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Intera 3.0T Quasar Dual ↗Class IIZ-2212-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- MR 5300Class IIZ-2213-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- MR 7700Class IIZ-2214-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- SmartPath to dStream for 3.0TClass IIZ-2215-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Z-2216-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- SmartPath to Ingenia Elition XClass IIZ-2217-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Upgrade to MR 7700Class IIZ-2218-2025 · Philips North America · posted Jul 30, 2025 · Open, Classified
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Z-2219-2025 · FUJIFILM Healthcare Americas Corporation · posted Jul 30, 2025 · Open, Classified
Devices had an unapproved slabbing software function enabled for use.
- Z-2220-2025 · Institut Georges Lopez · posted Jul 30, 2025 · Open, Classified
The potential for risk of microbiological contamination of products due to inability to ensure sterility assurance throughout aseptic manufacturing.
- Z-2221-2025 · Volcano Corp · posted Jul 30, 2025 · Open, Classified
Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating…
- Z-2189-2025 · Philips Medical Systems Nederland B.V. · posted Jul 29, 2025 · Open, Classified
Potential for temporary loss of imaging (X-ray) functionality due to software issue.
- BostonSight SCLERAL LensClass IIZ-2190-2025 · Boston Foundation For Sight · posted Jul 29, 2025 · Completed
Select orders of lenses were manufactured with errors impacting sagittal height, haptics, or both.
- BostonSight PROSE Lens.Class IIZ-2191-2025 · Boston Foundation For Sight · posted Jul 29, 2025 · Completed
Select orders of lenses were manufactured with errors impacting sagittal height, haptics, or both.
- Z-2192-2025 · Phasor Health, LLC · posted Jul 29, 2025 · Open, Classified
Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.
- Sprinter Cart/Sprinter Cart XL Infusion Pole. Component used to hold 2 x 2 kg infusion bags.Class IIZ-2193-2025 · Maquet Cardiopulmonary Gmbh · posted Jul 29, 2025 · Open, Classified
Device does not comply with the labeling requirements of international standard IEC 60601-1.
- Z-2194-2025 · Microbiologics Inc · posted Jul 29, 2025 · Open, Classified
KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.
- Z-2195-2025 · Microbiologics Inc · posted Jul 29, 2025 · Open, Classified
KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.
- Z-2196-2025 · LTS Therapy Systems, LLC · posted Jul 29, 2025 · Open, Classified
The nonwoven pads in the device do not absorb the saline solution or any other liquid intended to be used with the device.
- Z-2197-2025 · LTS Therapy Systems, LLC · posted Jul 29, 2025 · Open, Classified
The nonwoven pads in the device do not absorb the saline solution or any other liquid intended to be used with the device.
- Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide MonitorClass IIZ-2198-2025 · Capnia Inc · posted Jul 29, 2025 · Open, Classified
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher…
- Z-2188-2025 · Microvention Inc. · posted Jul 28, 2025 · Open, Classified
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other…
- Z-2172-2025 · Implant Direct Sybron Manufacturing LLC · posted Jul 25, 2025 · Open, Classified
The Drill stop kit contains incorrect components.
- Z-2174-2025 · HeartSine Technologies Ltd · posted Jul 25, 2025 · Open, Classified
Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)
- Z-2175-2025 · Remote Diagnostic Technologies Ltd. · posted Jul 25, 2025 · Open, Classified
Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed…
- Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting SystemsClass IIZ-2177-2025 · Steris Corporation · posted Jul 25, 2025 · Open, Classified
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
- Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting SystemsClass IIZ-2178-2025 · Steris Corporation · posted Jul 25, 2025 · Open, Classified
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
- Z-2149-2025 · Jinan Bodor Cnc Machine Co Ltd · posted Jul 23, 2025 · Open, Classified
The Bodor i Series products fail to meet certain applicable Federal standards, 21 CFR 1040.10(f)(1) and 21 CFR 1040.10(f)(2).
- Z-2162-2025 · Becton Dickinson & Co. · posted Jul 23, 2025 · Open, Classified
Potential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.
- Z-2163-2025 · Johnson & Johnson Vision Care, Inc. · posted Jul 23, 2025 · Open, Classified
Due to defects (bubbles/voids) identified during standard finished goods testing
- Z-2165-2025 · CooperVision, Inc. · posted Jul 23, 2025 · Open, Classified
Lens blisters may have an incomplete or leaking seal which may render them unsterile.
- Z-2166-2025 · CooperVision, Inc. · posted Jul 23, 2025 · Open, Classified
Lens blisters may have an incomplete or leaking seal which may render them unsterile.
- Z-2168-2025 · Medtronic, Inc. · posted Jul 23, 2025 · Open, Classified
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy…
- Z-2169-2025 · Medtronic, Inc. · posted Jul 23, 2025 · Open, Classified
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy…
- Stryker, StrykeFlow , REF: 0250070500, 2, Disposable Suction/Irrigator without Disposable TipClass IIZ-2170-2025 · Stryker Corporation · posted Jul 23, 2025 · Open, Classified
Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction.
- Z-2171-2025 · Stryker Corporation · posted Jul 23, 2025 · Open, Classified
Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction.
- Z-2157-2025 · Bridge to Life Ltd · posted Jul 22, 2025 · Open, Classified
The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit…
- Z-2139-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2140-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2141-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2142-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2143-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2144-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2145-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2146-2025 · Olympus Corporation of the Americas · posted Jul 21, 2025 · Open, Classified
Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
- Z-2147-2025 · Encore Medical, LP · posted Jul 21, 2025 · Open, Classified
Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. Issued discovered via a customer complaint.
- Z-2148-2025 · Encore Medical, LP · posted Jul 21, 2025 · Open, Classified
Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. Issued discovered via a customer complaint.
- Z-2151-2025 · Covidien · posted Jul 21, 2025 · Open, Classified
Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm…
- Estrone RIA, REF: DSL8700.Class IIZ-2152-2025 · Beckman Coulter Inc. · posted Jul 21, 2025 · Open, Classified
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for…
- Z-2153-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jul 21, 2025 · Open, Classified
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms…
- Z-2154-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jul 21, 2025 · Open, Classified
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms…
- Z-2155-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jul 21, 2025 · Open, Classified
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms…
- Z-2156-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jul 21, 2025 · Open, Classified
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms…
- Portable X-ray systemClass IIZ-2117-2025 · Digimed CO., Ltd · posted Jul 18, 2025 · Open, Classified
Some DIOX-602 units currently in use within the United States do not meet the minimum SSD requirement of 18 cm.
- SIGNA ArchitectClass IIZ-2130-2025 · GE Medical Systems, LLC · posted Jul 18, 2025 · Open, Classified
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit…
- SIGNA Architect AIR ↗Class IIZ-2131-2025 · GE Medical Systems, LLC · posted Jul 18, 2025 · Open, Classified
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit…
- Discovery MR750w 3.0TClass IIZ-2132-2025 · GE Medical Systems, LLC · posted Jul 18, 2025 · Open, Classified
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit…
- SIGNA PET/MRClass IIZ-2133-2025 · GE Medical Systems, LLC · posted Jul 18, 2025 · Open, Classified
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit…
- Z-2134-2025 · Philips Medical Systems Nederland B.V. · posted Jul 18, 2025 · Open, Classified
Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
- Z-2135-2025 · CooperSurgical, Inc. · posted Jul 18, 2025 · Open, Classified
The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
- Z-2136-2025 · CooperSurgical, Inc. · posted Jul 18, 2025 · Open, Classified
The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
- Z-2137-2025 · CooperSurgical, Inc. · posted Jul 18, 2025 · Open, Classified
The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
- Z-2138-2025 · Natus Neurology DBA Excel Tech., Ltd. (XLTEK) · posted Jul 18, 2025 · Open, Classified
Natus received two complaints of left side labeling show impedance for right side.
- Brand Name: NIO-AClass IIZ-2118-2025 · Waismed Ltd. · posted Jul 17, 2025 · Open, Classified
Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.
- Brand Name: NIO+Class IIZ-2119-2025 · Waismed Ltd. · posted Jul 17, 2025 · Open, Classified
Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.
- Z-2120-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jul 17, 2025 · Open, Classified
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the…
- Z-2121-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jul 17, 2025 · Open, Classified
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the…
- Z-2122-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jul 17, 2025 · Open, Classified
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the…
- Z-2123-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jul 17, 2025 · Open, Classified
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the…
- Z-2124-2025 · Microbiologics Inc · posted Jul 17, 2025 · Open, Classified
Affected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality control to fail and delayed diagnosis for the patient.
- Z-2125-2025 · Encore Medical, LP · posted Jul 17, 2025 · Open, Classified
Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. The use of this product may result in the incorrect product being selected. This could result in a revision surgery if used.
- Z-2126-2025 · Encore Medical, LP · posted Jul 17, 2025 · Open, Classified
Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. The use of this product may result in the incorrect product being selected. This could result in a revision surgery if used.
- Z-2127-2025 · Medtronic, Inc. · posted Jul 17, 2025 · Open, Classified
A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous…
- Z-2128-2025 · Medtronic, Inc. · posted Jul 17, 2025 · Open, Classified
A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous…
- Z-2111-2025 · Beyond Laser Systems, LLC · posted Jul 16, 2025 · Open, Classified
The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR…
- Z-2112-2025 · Beyond Laser Systems, LLC · posted Jul 16, 2025 · Open, Classified
The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR…
- Z-2113-2025 · Beyond Laser Systems, LLC · posted Jul 16, 2025 · Open, Classified
The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR…
- Z-2114-2025 · NeuroSync, Inc. · posted Jul 11, 2025 · Open, Classified
Eye syncing issues causing the device to be inoperable. Unable to be repaired due to discontinuation of support for the Pico 2 platform, which was no longer in production or serviceable as of August 31, 2023.
- Z-2115-2025 · Beckman Coulter, Inc. · posted Jul 11, 2025 · Completed
Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a delayed or improper diagnosis and/or…
- Z-2116-2025 · ICU Medical, Inc. · posted Jul 11, 2025 · Open, Classified
Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
- Z-2103-2025 · Boston Scientific Corporation · posted Jul 10, 2025 · Open, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due…
- Z-2104-2025 · Boston Scientific Corporation · posted Jul 10, 2025 · Open, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due…
- Z-2105-2025 · Boston Scientific Corporation · posted Jul 10, 2025 · Open, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due…
- Z-2106-2025 · Boston Scientific Corporation · posted Jul 10, 2025 · Open, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due…
- Z-2107-2025 · Preat Corp · posted Jul 10, 2025 · Open, Classified
Due to manufacturing error, digital analog rotational feature is 180 degrees off.
- Z-2108-2025 · Alcon Research LLC · posted Jul 10, 2025 · Open, Classified
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being…
- Z-2109-2025 · Alcon Research LLC · posted Jul 10, 2025 · Open, Classified
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being…
- Z-2110-2025 · Alcon Research LLC · posted Jul 10, 2025 · Open, Classified
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being…
- Brand Name: Hummingbird ↗Class IIZ-2100-2025 · Neurovision Medical Products Inc · posted Jul 9, 2025 · Open, Classified
Product packaging contains the wrong part number and kit contents.
- Z-2101-2025 · Medtronic Navigation, Inc.-Boxborough · posted Jul 9, 2025 · Open, Classified
A mechanical component that supports the O2 gantry to the support structure in affected devices may have been damaged during manufacture.
- Z-2097-2025 · Ge Medical Systems Israel, Functional Imaging · posted Jul 8, 2025 · Open, Classified
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
- Z-2098-2025 · Ge Medical Systems Israel, Functional Imaging · posted Jul 8, 2025 · Open, Classified
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
- Z-2099-2025 · Canary Medical, Inc. · posted Jul 8, 2025 · Open, Classified
Due to labeling error. Product incorrectly labeled.
- Z-2074-2025 · C.R. Bard Inc · posted Jul 3, 2025 · Open, Classified
Cather packaging may contain the incorrect French size.
- Z-2075-2025 · C.R. Bard Inc · posted Jul 3, 2025 · Open, Classified
Cather packaging may contain the incorrect French size.
- Z-2076-2025 · C.R. Bard Inc · posted Jul 3, 2025 · Open, Classified
Cather packaging may contain the incorrect French size.
- Z-2077-2025 · Geneoscopy, Inc. · posted Jul 3, 2025 · Open, Classified
Multiple plate failures documented as a result of the Low Positive Template Control being too high and out of range using the affected lot of ColoSense Test Kit.
- Z-2078-2025 · Becton, Dickinson and Company, BD Bio Sciences · posted Jul 3, 2025 · Open, Classified
Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
- Z-2079-2025 · Becton, Dickinson and Company, BD Bio Sciences · posted Jul 3, 2025 · Open, Classified
Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.
- Z-2086-2025 · Siemens Healthcare Diagnostics, Inc. · posted Jul 3, 2025 · Open, Classified
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
- Z-2087-2025 · Siemens Healthcare Diagnostics, Inc. · posted Jul 3, 2025 · Open, Classified
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
- Z-2088-2025 · Siemens Healthcare Diagnostics, Inc. · posted Jul 3, 2025 · Open, Classified
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
- Z-2089-2025 · Siemens Healthcare Diagnostics, Inc. · posted Jul 3, 2025 · Open, Classified
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
- Philips 3D9-3v Transducer.Class IIZ-2048-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips C5-1 Transducer.Class IIZ-2049-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips C5-2 Transducer.Class IIZ-2050-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips C9-5EC Transducer. ↗Class IIZ-2051-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips eL18-4 Transducer.Class IIZ-2052-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips eL18-4 EMT Transducer.Class IIZ-2053-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips L12-3 Transducer.Class IIZ-2054-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips L12-4 Transducer.Class IIZ-2055-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips L8-4 Transducer.Class IIZ-2056-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips L9-3 Transducer. ↗Class IIZ-2057-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips S5-1 Transducer.Class IIZ-2058-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips S7-2 Transducer.Class IIZ-2059-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips S7-3t Transducer.Class IIZ-2060-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips S8-3t Transducer.Class IIZ-2061-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips V6-2 Transducer.Class IIZ-2062-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips VL13-5 Transducer.Class IIZ-2063-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips X5-1 Transducer.Class IIZ-2064-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips X6-1 Transducer.Class IIZ-2065-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips X7-2t Transducer.Class IIZ-2066-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips X8-2t Transducer.Class IIZ-2067-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Philips XL14-3 Transducer.Class IIZ-2068-2025 · Philips Ultrasound, Inc · posted Jul 2, 2025 · Open, Classified
Ultrasound transducer devices were refurbished beyond their useful life.
- Z-2069-2025 · Corin Medical, Ltd. · posted Jul 2, 2025 · Open, Classified
Investigation has identified that the age of the nylon packaging material has caused a loss of integrity of some pouches.
- Z-2070-2025 · HemoCue AB · posted Jul 2, 2025 · Open, Classified
Glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not have been stored at the recommended labeled storage conditions which may have…
- Z-2071-2025 · Change Healthcare Canada Company · posted Jul 2, 2025 · Open, Classified
Due to complaints, software update may cause software to unexpectedly shutdown.
- Z-2072-2025 · Change Healthcare Canada Company · posted Jul 2, 2025 · Open, Classified
Due to complaints, software update may cause software to unexpectedly shutdown.
- Z-2073-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jul 2, 2025 · Open, Classified
Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO…
Class III recalls (5)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-2090-2025 · Stratasys Ltd · posted Jul 7, 2025 · Completed
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
- Z-2091-2025 · Stratasys Ltd · posted Jul 7, 2025 · Completed
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
- Z-2092-2025 · Stratasys Ltd · posted Jul 7, 2025 · Completed
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
- Z-2093-2025 · Stratasys Ltd · posted Jul 7, 2025 · Completed
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
- Z-2094-2025 · Stratasys Ltd · posted Jul 7, 2025 · Completed
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.