BROSELOW recallZ-1904-2025 · SunMed Holdings, LLC · initiated May 15, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from SunMed Holdings, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.
FDA root cause category: Process control
What the company is doing
On May 15, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were given detailed information about the incorrect information printed on the tape. Customers were instructed to identify and quarantine all affected product in inventory. Customers should destroy and dispose of all affected product. If product was further distributed, distributors should identify their customers/consignees and notify them of the recall. The firm will provide replacement products once customer confirms destruction of affected product.
Recall details
- Recall number
- Z-1904-2025
- Status
- Open, Classified
- Initiated by the company
- May 15, 2025
- Posted by the FDA
- Jun 20, 2025
- Recalling firm
- SunMed Holdings, LLC
- Quantity
- 9900 eaches
- Product code
- OKI
- Distribution
- US distribution nationwide. International distribution to Canada.
Product description
Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print Version The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.
Code information (models, lots, serial numbers)
UDI-DI EA: 10889483588970 CS: 30889483588974 Lot 0004316661
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BROSELOW™ 7700REAAirlife · Emergency Response Safety Kit2 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page