Skip to content

BROSELOW recallZ-1904-2025 · SunMed Holdings, LLC · initiated May 15, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from SunMed Holdings, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.

FDA root cause category: Process control

What the company is doing

On May 15, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were given detailed information about the incorrect information printed on the tape. Customers were instructed to identify and quarantine all affected product in inventory. Customers should destroy and dispose of all affected product. If product was further distributed, distributors should identify their customers/consignees and notify them of the recall. The firm will provide replacement products once customer confirms destruction of affected product.

Recall details

Recall number
Z-1904-2025
Status
Open, Classified
Initiated by the company
May 15, 2025
Posted by the FDA
Jun 20, 2025
Recalling firm
SunMed Holdings, LLC
Quantity
9900 eaches
Product code
OKI
Distribution
US distribution nationwide. International distribution to Canada.

Product description

Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print Version The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.

Code information (models, lots, serial numbers)

UDI-DI EA: 10889483588970 CS: 30889483588974 Lot 0004316661