Skip to content

MicroPlex Helical Soft recallZ-2188-2025 · Microvention Inc. · initiated May 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microvention Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.

FDA root cause category: Device Design

What the company is doing

On 5/29/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inform all individuals within your organization of the recall and forward the recall notice to any organizations that may have received the affected products. 2) Complete and return the customer acknowledgement form via email: [email protected] 3) If your institution has affected inventory call firm's Customer Care at 800-888-3786 or email [email protected] Please direct questions to the firm at [email protected]

Recall details

Recall number
Z-2188-2025
Status
Open, Classified
Initiated by the company
May 29, 2025
Posted by the FDA
Jul 28, 2025
Recalling firm
Microvention Inc.
Quantity
1451
Product code
KRD
FDA submissions
K150958
Distribution
OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France, Kazakhstan, Slovakia, Norway, Greece, Serbia, Turkey, Czech Republic, United Kingdom, Netherlands, Russian Federation, Kuwait, Republic of North Macedonia, Switzerland, Saudi Arabia, Georgia, Lithuania, Romania, Croatia, Chile, Hong Kong , India, Mexico, Taiwan, Vietnam

Product description

LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400

Code information (models, lots, serial numbers)

REF/GTIN/Lot: 8LP2S20/(01)00812636020105/0000177644, 0000180340, 0000223990, 0000235256, 0000254143, 0000254142, 0000272315, 0000272316, 0000281719, 0000294361, 0000295741, 0000296446, 0000751662, 0000751661, 0001026153; 8LP2S400/(01)00812636020112/0000199869, 0000225205, 0000708506, 0000863001, 0001026156

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates