Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number:… recallZ-1939-2025 · Stryker GmbH · initiated May 22, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
FDA root cause category: Under Investigation by firm
What the company is doing
Stryker notified consignees on about 05/22/2025 via letter. Consignees were instructed to immediately check their inventory for affected units and quarantine any on hand, and complete and return the Business Reply Form by email to . Upon receipt of the completed Business Reply Form, Stryker will contact customers to arrange for the return of their product(s). If consignees have further distributed the product, they were instructed to notify their customers.
Recall details
- Recall number
- Z-1939-2025
- Status
- Open, Classified
- Initiated by the company
- May 22, 2025
- Posted by the FDA
- Jun 10, 2025
- Recalling firm
- Stryker GmbH
- Quantity
- 1
- Product code
- HSB
- FDA submissions
- K101438
- Distribution
- Worldwide - US Nationwide distribution in the states of California and Missouri. The countries of Belgium, Japan, Netherlands.
Product description
Femoral Nail, LEFT T2 GTN 8x300 mm. Catalog Number: 1850-0830S.
Code information (models, lots, serial numbers)
Catalog Number: 1850-0830S. GTIN: 04546540646293. Batch lot number: K09F126
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- T2 1850-0830SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 96924).
- Z-1940-2025 · Open, ClassifiedFemoral Nail, RIGHT T2 GTN 8x300 mm. Catalog Number: 1851-0830S.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page