VITROS recallZ-2038-2025 · Ortho-Clinical Diagnostics, Inc. · initiated May 9, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
FDA root cause category: Software design
What the company is doing
On May 22, 2025, "URGENT PRODUCT CORRECTION NOTIFICATION" letters were sent to customers. REQUIRED ACTIONS - To prevent this issue from occurring, the shelf expiration date must be entered when manually loading VITROS Diluent Packs on VITROS 4600/5600/XT 7600 Systems. - View the Reagent Management Supply 3 screen to check for any onboard VITROS Diluent Packs marked as expired. o For all VITROS Diluent Packs marked as expired, navigate to View By Reagent View Dil/ANC and confirm the shelf expiration date matches the shelf expiration date on the product labeling. o If the shelf expiration dates do not match, discard the VITROS Diluent Pack and load a new VITROS Diluent Pack. QuidelOrtho will credit your discarded inventory. (Credit will only be issued for within-expiry products.) - Complete and return the enclosed Confirmation of Receipt form no later than May 30, 2025. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. - Save this notification with your User Documentation or post this notification by each VITROS 4600/5600/XT 7600 System in your laboratory until the issue has been resolved. - Please forward this notification if the affected product was distributed outside of your facility. - If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Resolution QuidelOrtho has determined root cause, and this issue will be resolved in an upcoming software update. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Recall details
- Recall number
- Z-2038-2025
- Status
- Open, Classified
- Initiated by the company
- May 9, 2025
- Posted by the FDA
- Jun 27, 2025
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 3117 units (813 US, 2304 OUS)
- Product code
- JJE
- FDA submissions
- K081543
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.
Product description
Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
Code information (models, lots, serial numbers)
VITROS 5600 Integrated System; Product Code: 6802413; UDI: 10758750002740; VITROS 5600 Integrated System - Certified/Refurbished; Product Code: 6802915; UDI: 10758750002740; Range of installed serial numbers: 56000143-56005096 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802915 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6802413Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE12 recalls69 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96946).
- Z-2039-2025 · Open, ClassifiedBrand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software Version: version 3.2 to version 3.8.3 (all available so…
- Z-2040-2025 · Open, ClassifiedBrand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page