EPIONE Model 30-0001 recallZ-1927-2025 · Quantum Surgical Sas · initiated May 23, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Quantum Surgical Sas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user.

FDA root cause category: Under Investigation by firm

What the company is doing

On or about 05/23/2025, the firm sent via emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that due to a software issue the EPIONE device can place the instrument guide at an incorrect position, compared to the planning defined by the user. Customers are instructed to: 1. Pending the implementation of a corrective action, Quantum Surgical asks all EPIONE device users to NOT plan trajectories along a strictly horizontal direction. To determine if a trajectory is strictly horizontal the following methods can be used: . Patient images acquired in prone or supine orientation: o If the entry and target point are on the same native (non-oblique) axial plane and are on the same native (non-oblique) coronal plane -> the trajectory is strictly horizontal. o If the trajectory is perpendicular to the sagittal plane -> the trajectory is strictly horizontal. . Patient images acquired in lateral decubitus orientation: o If the entry and target point are on the same native (non-oblique) axial plane and are on the same native (non-oblique) sagittal plane -> the trajectory is strictly horizontal. o If the trajectory is perpendicular to the coronal plane -> the trajectory is strictly horizontal. Note - Quantum Surgical reminds that the instrument insertion shall be initiated only after the "Robot on Target" message is displayed, as mentioned in Epione User Manual on page 100. For question or assistance, contact Quantum Surgical on +33 (0) 4 48 19 40 50 between 9am and 6pm, Monday to Friday.

Recall details

Recall number
Z-1927-2025
Status
Open, Classified
Initiated by the company
May 23, 2025
Posted by the FDA
Jun 9, 2025
Recalling firm
Quantum Surgical Sas
Quantity
4 units in U.S.
Product code
JAK
FDA submissions
K223758
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of FL,IL, IN, and TX. The countries of France, Germany and United Kingdom.

Product description

EPIONE Model 30-0001 UDI-DI code: 03760305400031 The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous ablation procedures in the abdomen, performed by physicians trained for CT procedures and performed under general anaesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G.

Code information (models, lots, serial numbers)

Model Number: 30-0001 UDI-DI code: 03760305400031 Software Version: 1.0.6.1 Serial Numbers (U.S.): 0322002, 0323006, 0323003 and 0324001