KWIK-STIK recallZ-1957-2025 · Microbiologics Inc · initiated May 27, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.

FDA root cause category: Mixed-up of materials/components

What the company is doing

On May 27, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed of the affected products containing the wrong microorganisms. Customers were instructed to review lab procedures to understand how this information affects their usage, and then to use or discard the product depending on the lab procedures and how the issue affects usage. Distributors of the affected product were instructed to contact end users who have received the affected items.

Recall details

Recall number
Z-1957-2025
Status
Open, Classified
Initiated by the company
May 27, 2025
Posted by the FDA
Jun 11, 2025
Recalling firm
Microbiologics Inc
Quantity
3 total
Product code
JTR
FDA submissions
K861022
Distribution
US Nationwide distribution in the states of California and Pennsylvania and the country of New Zealand.

Product description

KWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived from ATCC 17960; 0621K, Staphylococcus aureus derived from NCTC 12493

Code information (models, lots, serial numbers)

01065K: UDI-DI 30845357032868, lot 1065-25-51; 0621K: UDI-DI 30845357014666, lot 621-73-21