Spectral CT. Computed tomography X-ray system recallZ-2034-2025 · Philips North America Llc · initiated May 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.

FDA root cause category: Under Investigation by firm

What the company is doing

An "URGENT Medical Device Correction" notice dated 5/21/25 was mailed to consignees. The notification instructs consignees on how to identify if their devices have affected software downloaded. Consignees may continue to use affected systems in accordance with the intended use and recommended actions provided in the recall notification in Appendix A. Appendix A summarizes the ten software issues and the actions recommended by the manufacturer that the consignee follow. Consignees are to complete and return the provided response form, forward the recall notification to users, and retain a copy with all affected systems until a solution is installed. Philips will contact consignees to schedule time for a Field Service Engineer to install the solution and resolve the issue (Reference FCO72800828/FCO 72800829). Consignees with any questions can contact Customer Care Solutions at 1-800-722-9377.

Recall details

Recall number
Z-2034-2025
Status
Open, Classified
Initiated by the company
May 29, 2025
Posted by the FDA
Jun 26, 2025
Recalling firm
Philips North America Llc
Quantity
370 units
Product code
JAK
FDA submissions
K203020
Distribution
Domestic: AZ, CA, CO, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV; Foreign: Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Lithuania, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.

Product description

Spectral CT. Computed tomography X-ray system.

Code information (models, lots, serial numbers)

Show the full code information (11k characters)

Part No. 728333; UDI: (01)00884838101111(21)10014, (01)00884838101111(21)10015, (01)00884838101111(21)10016, (01)00884838101111(21)10017, (01)00884838101111(21)10018, (01)00884838101111(21)10019, (01)00884838101111(21)10020, (01)00884838101111(21)10021, (01)00884838101111(21)10022, (01)00884838101111(21)10023, (01)00884838101111(21)10024, (01)00884838101111(21)10025, (01)00884838101111(21)10027, (01)00884838101111(21)10031, (01)00884838101111(21)10032, (01)00884838101111(21)10033, (01)00884838101111(21)10034, (01)00884838101111(21)10035, (01)00884838101111(21)10036, (01)00884838101111(21)10038, (01)00884838101111(21)10039, (01)00884838101111(21)10040, (01)00884838101111(21)10041, (01)00884838101111(21)10042, (01)00884838101111(21)10043, (01)00884838101111(21)10044, (01)00884838101111(21)10045, (01)00884838101111(21)10046, (01)00884838101111(21)10047, (01)00884838101111(21)10048, (01)00884838101111(21)10049, (01)00884838101111(21)10050, (01)00884838101111(21)10052, (01)00884838101111(21)10053, (01)00884838101111(21)10055, (01)00884838101111(21)10056, (01)00884838101111(21)10057, (01)00884838101111(21)10058, (01)00884838101111(21)10059, (01)00884838101111(21)10060, (01)00884838101111(21)10061, (01)00884838101111(21)10062, (01)00884838101111(21)10063, (01)00884838101111(21)10065, (01)00884838101111(21)10066, (01)00884838101111(21)10067, (01)00884838101111(21)10068, (01)00884838101111(21)10069, (01)00884838101111(21)10070, (01)00884838101111(21)10071, (01)00884838101111(21)10072, (01)00884838101111(21)10074, (01)00884838101111(21)10075, (01)00884838101111(21)10076, (01)00884838101111(21)10077, (01)00884838101111(21)10078, (01)00884838101111(21)10079, (01)00884838101111(21)10080, (01)00884838101111(21)10081, (01)00884838101111(21)10082, (01)00884838101111(21)10084, (01)00884838101111(21)10085, (01)00884838101111(21)10086, (01)00884838101111(21)10087, (01)00884838101111(21)10088, (01)00884838101111(21)10089, (01)00884838101111(21)10090, (01)00884838101111(21)10091, (01)00884838101111(21)10092, (01)00884838101111(21)10093, (01)00884838101111(21)10094, (01)00884838101111(21)10095, (01)00884838101111(21)10096, (01)00884838101111(21)10097, (01)00884838101111(21)10098, (01)00884838101111(21)10099, (01)00884838101111(21)10100, (01)00884838101111(21)10101, (01)00884838101111(21)10102, (01)00884838101111(21)10103, (01)00884838101111(21)10104, (01)00884838101111(21)10105, (01)00884838101111(21)10106, (01)00884838101111(21)10107, (01)00884838101111(21)10108, (01)00884838101111(21)10109, (01)00884838101111(21)10110, (01)00884838101111(21)10111, (01)00884838101111(21)10112, (01)00884838101111(21)10113, (01)00884838101111(21)10114, (01)00884838101111(21)10115, (01)00884838101111(21)10116, (01)00884838101111(21)10117, (01)00884838101111(21)10118, (01)00884838101111(21)10119, (01)00884838101111(21)10120, (01)00884838101111(21)10121, (01)00884838101111(21)10122, (01)00884838101111(21)10123, (01)00884838101111(21)10124, (01)00884838101111(21)10125, (01)00884838101111(21)10126, (01)00884838101111(21)10127, (01)00884838101111(21)10128, (01)00884838101111(21)10129, (01)00884838101111(21)10130, (01)00884838101111(21)10131, (01)00884838101111(21)10132, (01)00884838101111(21)10133, (01)00884838101111(21)10134, (01)00884838101111(21)10135, (01)00884838101111(21)10136, (01)00884838101111(21)10137, (01)00884838101111(21)10138, (01)00884838101111(21)10139, (01)00884838101111(21)10140, (01)00884838101111(21)10141, (01)00884838101111(21)10142, (01)00884838101111(21)10143, (01)00884838101111(21)10144, (01)00884838101111(21)10145, (01)00884838101111(21)10146, (01)00884838101111(21)10147, (01)00884838101111(21)10148, (01)00884838101111(21)10149, (01)00884838101111(21)10150, (01)00884838101111(21)10151, (01)00884838101111(21)10152, (01)00884838101111(21)10153, (01)00884838101111(21)10154, (01)00884838101111(21)10155, (01)00884838101111(21)10156, (01)00884838101111(21)10157, (01)00884838101111(21)10158, (01)00884838101111(21)10159, (01)00884838101111(21)10160, (01)00884838101111(21)10161, (01)00884838101111(21)10162, (01)00884838101111(21)10163, (01)00884838101111(21)10164, (01)00884838101111(21)10165, (01)00884838101111(21)10166, (01)00884838101111(21)10167, (01)00884838101111(21)10168, (01)00884838101111(21)10169, (01)00884838101111(21)10170, (01)00884838101111(21)10171, (01)00884838101111(21)10172, (01)00884838101111(21)10173, (01)00884838101111(21)10174, (01)00884838101111(21)10175, (01)00884838101111(21)10176, (01)00884838101111(21)10177, (01)00884838101111(21)10178, (01)00884838101111(21)10179, (01)00884838101111(21)10180, (01)00884838101111(21)10181, (01)00884838101111(21)10182, (01)00884838101111(21)10183, (01)00884838101111(21)10184, (01)00884838101111(21)10185, (01)00884838101111(21)10186, (01)00884838101111(21)10187, (01)00884838101111(21)10188, (01)00884838101111(21)10189, (01)00884838101111(21)10190, (01)00884838101111(21)10191, (01)00884838101111(21)10192, (01)00884838101111(21)10193, (01)00884838101111(21)10194, (01)00884838101111(21)10195, (01)00884838101111(21)10196, (01)00884838101111(21)10197, (01)00884838101111(21)10198, (01)00884838101111(21)10199, (01)00884838101111(21)10200, (01)00884838101111(21)10201, (01)00884838101111(21)10202, (01)00884838101111(21)10203, (01)00884838101111(21)10204, (01)00884838101111(21)10205, (01)00884838101111(21)10206, (01)00884838101111(21)10207, (01)00884838101111(21)10208, (01)00884838101111(21)10209, (01)00884838101111(21)10210, (01)00884838101111(21)10211, (01)00884838101111(21)10212, (01)00884838101111(21)10213, (01)00884838101111(21)10214, (01)00884838101111(21)10215, (01)00884838101111(21)10217, (01)00884838101111(21)10218, (01)00884838101111(21)10219, (01)00884838101111(21)10220, (01)00884838101111(21)10221, (01)00884838101111(21)10222, (01)00884838101111(21)10223, (01)00884838101111(21)10224, (01)00884838101111(21)10225, (01)00884838101111(21)10226, (01)00884838101111(21)10227, (01)00884838101111(21)10228, (01)00884838101111(21)10229, (01)00884838101111(21)10230, (01)00884838101111(21)10231, (01)00884838101111(21)10232, (01)00884838101111(21)10233, (01)00884838101111(21)10234, (01)00884838101111(21)10235, (01)00884838101111(21)10236, (01)00884838101111(21)10237, (01)00884838101111(21)10238, (01)00884838101111(21)10239, (01)00884838101111(21)10240, (01)00884838101111(21)10241, (01)00884838101111(21)10242, (01)00884838101111(21)10243, (01)00884838101111(21)10244, (01)00884838101111(21)10245, (01)00884838101111(21)10246, (01)00884838101111(21)10247, (01)00884838101111(21)10248, (01)00884838101111(21)10249, (01)00884838101111(21)10250, (01)00884838101111(21)10251, (01)00884838101111(21)10252, (01)00884838101111(21)10253, (01)00884838101111(21)10254, (01)00884838101111(21)10255, (01)00884838101111(21)10256, (01)00884838101111(21)10257, (01)00884838101111(21)10258, (01)00884838101111(21)10259, (01)00884838101111(21)10260, (01)00884838101111(21)10261, (01)00884838101111(21)10262, (01)00884838101111(21)10263, (01)00884838101111(21)10264, (01)00884838101111(21)10265, (01)00884838101111(21)10266, (01)00884838101111(21)10267, (01)00884838101111(21)10268, (01)00884838101111(21)10269, (01)00884838101111(21)10270, (01)00884838101111(21)10271, (01)00884838101111(21)10272, (01)00884838101111(21)10273, (01)00884838101111(21)10274, (01)00884838101111(21)10275, (01)00884838101111(21)10276, (01)00884838101111(21)10277, (01)00884838101111(21)10278, (01)00884838101111(21)10279, (01)00884838101111(21)10280, (01)00884838101111(21)10281, (01)00884838101111(21)10284, (01)00884838101111(21)10285, (01)00884838101111(21)10286, (01)00884838101111(21)10287, (01)00884838101111(21)10288, (01)00884838101111(21)10289, (01)00884838101111(21)10290, (01)00884838101111(21)10291, (01)00884838101111(21)10292, (01)00884838101111(21)10293, (01)00884838101111(21)10294, (01)00884838101111(21)10298, (01)00884838101111(21)10299, (01)00884838101111(21)10300, (01)00884838101111(21)10301, (01)00884838101111(21)12000, (01)00884838101111(21)12001, (01)00884838101111(21)12002, (01)00884838101111(21)12003, (01)00884838101111(21)12004, (01)00884838101111(21)12005, (01)00884838101111(21)12006, (01)00884838101111(21)12007, (01)00884838101111(21)12008, (01)00884838101111(21)12009, (01)00884838101111(21)396001, (01)00884838101111(21)396003, (01)00884838101111(21)396004, (01)00884838101111(21)396005, (01)00884838101111(21)396006, (01)00884838101111(21)396007, (01)00884838101111(21)396008, (01)00884838101111(21)396009, (01)00884838101111(21)396010, (01)00884838101111(21)396011; Serial No. 10013, 10014, 10015, 10016, 10017, 10018, 10019, 10020, 10021, 10022, 10023, 10024, 10025, 10027, 10031, 10032, 10033, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10042, 10043, 10044, 10045, 10046, 10047, 10048, 10049, 10050, 10052, 10053, 10055, 10056, 10057, 10058, 10059, 10060, 10061, 10062, 10063, 10065, 10066, 10067, 10068, 10069, 10070, 10071, 10072, 10074, 10075, 10076, 10077, 10078, 10079, 10080, 10081, 10082, 10084, 10085, 10086, 10087, 10088, 10089, 10090, 10091, 10092, 10093, 10094, 10095, 10096, 10097, 10098, 10099, 10100, 10101, 10102, 10103, 10104, 10105, 10106, 10107, 10108, 10109, 10110, 10111, 10112, 10113, 10114, 10115, 10116, 10117, 10118, 10119, 10120, 10121, 10122, 10123, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10134, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10145, 10146, 10147, 10148, 10149, 10150, 10151, 10152, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162, 10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10189, 10190, 10191, 10192, 10193, 10194, 10195, 10196, 10197, 10198, 10199, 10200, 10201, 10202, 10203, 10204, 10205, 10206, 10207, 10208, 10209, 10210, 10211, 10212, 10213, 10214, 10215, 10217, 10218, 10219, 10220, 10221, 10222, 10223, 10224, 10225, 10226, 10227, 10228, 10229, 10230, 10231, 10232, 10233, 10234, 10235, 10236, 10237, 10238, 10239, 10240, 10241, 10242, 10243, 10244, 10245, 10246, 10247, 10248, 10249, 10250, 10251, 10252, 10253, 10254, 10255, 10256, 10257, 10258, 10259, 10260, 10261, 10262, 10263, 10264, 10265, 10266, 10267, 10268, 10269, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10279, 10280, 10281, 10284, 10285, 10286, 10287, 10288, 10289, 10290, 10291, 10292, 10293, 10294, 10298, 10299, 10300, 10301, 12000, 12001, 12002, 12003, 12004, 12005, 12006, 12007, 12008, 12009, 396001, 396003, 396004, 396005, 396006, 396007, 396008, 396009, 396010, 396011.