Skip to content

Access Vitamin B12 Reagent, Catalog Number 33000, Lot… recallZ-2115-2025 · Beckman Coulter, Inc. · initiated Feb 27, 2025

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a delayed or improper diagnosis and/or repeat testing.

FDA root cause category: Process control

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 2/19/25 was sent to customers. ACTION: Discontinue using Access Vitamin B12 reagent lot 439850. Discard all remaining reagent packs from this lot following your laboratory procedures. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient results. Contact your local Beckman Coulter representative for replacement product requests. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com You may request replacement product by contacting your local Beckman Coulter Representative for replacement.

Recall details

Recall number
Z-2115-2025
Status
Completed
Initiated by the company
Feb 27, 2025
Posted by the FDA
Jul 11, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
10,000 units
Product code
CDD
FDA submissions
K955436
Distribution
International distribution to the country of Andorra, Bulgaria, Denmark, Estonia, France, Germany, Greece, Ireland, Italy, Maldives, Netherlands, Portugal, Romania, Spain, Switzerland, Turkey, United Kingdom of Great Britain and Northern Ireland

Product description

Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850

Code information (models, lots, serial numbers)

UDI-DI: 15099590224301; Lot Number: 439850

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page