Skip to content

Servo-u MR Ventilator System. Model Number: 6888800 recallZ-2015-2025 · Maquet Critical Care AB · initiated May 30, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Critical Care AB or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

FDA root cause category: Under Investigation by firm

What the company is doing

Maquet Critical Care AB, a subsidiary of Getinge, notified consignees on about 05/30/2025 via FedEx letter. Consignees were instructed to configure all affected Servo ventilators with "Extended leakage test" set to "Off", per chapter 8.2.4 Configuration in the User's manuals and all users are recommended to follow the updated instructions for use of the User's manuals chapter 9.5.1 Patient circuit test, which is provided. Consignees are instructed to ensure all devices in inventory have been properly addressed as described above, to ensure the message is forwarded to all affected personnel and to individuals if units were transferred. Additionally, consignees were requested to complete and return the provided response form.

Recall details

Recall number
Z-2015-2025
Status
Open, Classified
Initiated by the company
May 30, 2025
Posted by the FDA
Jun 27, 2025
Recalling firm
Maquet Critical Care AB
Quantity
129 units
Product code
CBK
FDA submissions
K201874
Distribution
US Nationwide distribution and Puerto Rico. No international distribution.

Product description

Servo-u MR Ventilator System. Model Number: 6888800.

Code information (models, lots, serial numbers)

Model Number: 6888800. UDI-DI: 07325710010518. All software versions.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K201874 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97032).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page