VITROS recallZ-2073-2025 · Ortho-Clinical Diagnostics, Inc. · initiated Jun 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.

FDA root cause category: Under Investigation by firm

What the company is doing

Quidel Ortho notified consignees on about 06/12/2025 via FedEx letter. Distributors were instructed to discontinue distribution and discard inventory of any affected lots, send the provided customer letter and Confirmation of Receipt to any affected customers, and complete and return their Confirmation of Receipt. Customers were instructed that if their laboratory is unable to generate QC results within acceptable range using lots from GEN 15 of VITROS OP Reagent, discontinue using, render unusable, and discard your inventory of VITROS OP Reagent, GEN 15. Additionally, there were requested to complete and return the Confirmation of Receipt, maintain the notification with all records, and provide the notification if affected units were further distributed.

Recall details

Recall number
Z-2073-2025
Status
Open, Classified
Initiated by the company
Jun 12, 2025
Posted by the FDA
Jul 2, 2025
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
2,017 units
Product code
DJG
FDA submissions
K062460
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Product description

VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.

Code information (models, lots, serial numbers)

Model Number: 680 1997. UDI-DI: 10758750001835. Lot Numbers: 1527-15-2749 (exp. 17-Nov-2025), 1527-15-2771 (exp. 17-Dec-2025), 1527-15-2818 (exp. 11-Apr-2026), 1527-15-2871 (exp. 15-May-2026).