VITROS recallZ-2073-2025 · Ortho-Clinical Diagnostics, Inc. · initiated Jun 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.
FDA root cause category: Under Investigation by firm
What the company is doing
Quidel Ortho notified consignees on about 06/12/2025 via FedEx letter. Distributors were instructed to discontinue distribution and discard inventory of any affected lots, send the provided customer letter and Confirmation of Receipt to any affected customers, and complete and return their Confirmation of Receipt. Customers were instructed that if their laboratory is unable to generate QC results within acceptable range using lots from GEN 15 of VITROS OP Reagent, discontinue using, render unusable, and discard your inventory of VITROS OP Reagent, GEN 15. Additionally, there were requested to complete and return the Confirmation of Receipt, maintain the notification with all records, and provide the notification if affected units were further distributed.
Recall details
- Recall number
- Z-2073-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 12, 2025
- Posted by the FDA
- Jul 2, 2025
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 2,017 units
- Product code
- DJG
- FDA submissions
- K062460
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Product description
VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.
Code information (models, lots, serial numbers)
Model Number: 680 1997. UDI-DI: 10758750001835. Lot Numbers: 1527-15-2749 (exp. 17-Nov-2025), 1527-15-2771 (exp. 17-Dec-2025), 1527-15-2818 (exp. 11-Apr-2026), 1527-15-2871 (exp. 15-May-2026).
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6801997Ortho-Clinical Diagnostics, Inc. · ENZYME IMMUNOASSAY, OPIATES1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.