Skip to content

ADVIA 120/2120/2120i SETpoint Calibrator recallZ-2086-2025 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.

FDA root cause category: Under Investigation by firm

What the company is doing

An Urgent Medical Device Correction notification dated 6/4/25 was mailed to consignees. This notification provides consignees with calibration instructions given the issue found in affected lots. Additionally, instructions are given on how to proceed with product depending on performance. Consignees are to discontinue use of and discard of the calibrator SETPOINT lots impacted by this recall. Consignee locations are to review the provided notification with their Medical Directors to determine the appropriate course of action. A copy of the recall notification should be retained in laboratory records and shared with users. Consignees are to complete and return the provided Effectiveness Check Form within 30 days of receipt of the recall notification. Consignees with any questions can contact their Siemens Customer Care Center or local Siemens Healthineers technical support representative.

Recall details

Recall number
Z-2086-2025
Status
Open, Classified
Initiated by the company
Jun 4, 2025
Posted by the FDA
Jul 3, 2025
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
2,020 units
Product code
KRZ
FDA submissions
K840261
Distribution
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.

Product description

ADVIA 120/2120/2120i SETpoint Calibrator

Code information (models, lots, serial numbers)

Siemens Material Number 10312285; UDI: 00630414224343, 00630414224343; Lot No. SP251033, SP251043.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97037).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page