Skip to content

TrueDent Cyan, TDM 120, Model No. OBJ-09165 recallZ-2091-2025 · Stratasys Ltd · initiated Jan 19, 2024

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stratasys Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.

FDA root cause category: Software design

What the company is doing

The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.

Recall details

Recall number
Z-2091-2025
Status
Completed
Initiated by the company
Jan 19, 2024
Posted by the FDA
Jul 7, 2025
Recalling firm
Stratasys Ltd
Quantity
100
Product code
EBI
FDA submissions
K220771
Distribution
US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.

Product description

TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliances

Code information (models, lots, serial numbers)

UDI-DI 7290018915400 Lots 23485 (exp. 08-Nov-2024) and 22904 (exp. 09-Feb-2024)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

4 devices cleared under K220771 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97021).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page