Skip to content

StrykeFlow recallZ-2170-2025 · Stryker Corporation · initiated Jun 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction.

FDA root cause category: Under Investigation by firm

What the company is doing

On June 18, 2025, Stryker issued a "Urgent: Medical Device Recall" Notification to affected consignees. On November 20, 2025, Stryker initiated a recall notification update. Stryker ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product. 3. If affected product is found, quarantine and initiate the return process for affected units. 3a. Please complete Business Reply Form or use QR Code on Attachment A to return product. 4. If affected product is NOT found: 4a. Please complete Business Reply Form or use QR Code on Attachment A. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Response is required.

Recall details

Recall number
Z-2170-2025
Status
Open, Classified
Initiated by the company
Jun 18, 2025
Posted by the FDA
Jul 23, 2025
Recalling firm
Stryker Corporation
Product code
GCJ
FDA submissions
K934094
Distribution
Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Argentina, Philippines, Singapore, China, Japan, Canada, Korea.

Product description

Stryker, StrykeFlow , REF: 0250070500, 2, Disposable Suction/Irrigator without Disposable Tip

Code information (models, lots, serial numbers)

UDI: 07613327061390

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

7 devices cleared under K934094 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97039).

Sources and updates