Baxter SIGMA Spectrum Infusion Pump V6, Model Number… recallZ-2095-2025 · Baxter Healthcare Corporation · initiated Jun 20, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain Spectrum infusion pumps may have an incorrect version of software.

FDA root cause category: Employee error

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 6/20/25 was sent to customers. Actions to be Taken by Customers 1. Immediately locate Spectrum pumps with the affected serial numbers and remove them from service. The product code and serial number can be found on the bottom of the infusion pump. 2. Check and record the software version installed on the pump. The software on a V6 pump should begin with the number 6 and the software on a V8 pump should begin with the number 8. The software version is located on the startup screen upon powering up the pump as pictured on the next page. Additionally, the software version is indicated in the pump menu which can be accessed by following the steps. For V6: From the Pump Information screen, press the sw info soft key to display the software version screen. For V8: From the Select Care Area screen, press the options menu soft key. Select User Options and press OK. Select View Information and press OK. Select Pump Information and press OK to display the Pump Info screen. 3. Contact Baxter to confirm the software version on the affected pumps and to schedule service if required. Baxter Global Technical Services can be reached at 800-843-7867 (select option 1, then option 2), Monday through Friday, between 7:00 am and 7:00 pm Eastern Time. 4. If you received this communication directly from Baxter, acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them

Recall details

Recall number
Z-2095-2025
Status
Open, Classified
Initiated by the company
Jun 20, 2025
Posted by the FDA
Jul 17, 2025
Recalling firm
Baxter Healthcare Corporation
Quantity
14 units
Product code
PHC
FDA submissions
K133801
Distribution
US States: CA, FL, IL, KS, MI, MS, NY, OH, OK, and VA.

Product description

Baxter SIGMA Spectrum Infusion Pump V6, Model Number 35700BAX

Code information (models, lots, serial numbers)

UDI-DI: 00085412091570; Serial Numbers: 765100 , 772744 , 789742 , 841197 , 867753 , 920606 , 1024015 770164 788724 , 793079 , 842498 , 918342 , 962810 , 1027523

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97072).