KWIK-STIK and LYFO DISK, quality control kit for culture… recallZ-2124-2025 · Microbiologics Inc · initiated Jun 17, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality control to fail and delayed diagnosis for the patient.

FDA root cause category: Under Investigation by firm

What the company is doing

On June 17. 2025, the firm began notifying affected customers via "Urgent Medical Device Recall" letters delivered through email. Customers were instructed to review lab procedures to understand how the information may affect usage; use or discard product depending on lab procedures. Complete and return the response form to [email protected]. If product was further distributed or transferred to other facilities or customers, they should also be notified. Customers may contact Microbiologics if a replacement kit is needed. If you have any questions or concerns, contact Recall Support team at 320.229.7080 or [email protected]. Collect calls may be made.

Recall details

Recall number
Z-2124-2025
Status
Open, Classified
Initiated by the company
Jun 17, 2025
Posted by the FDA
Jul 17, 2025
Recalling firm
Microbiologics Inc
Quantity
21 total
Product code
JTR
FDA submissions
K861022
Distribution
US distribution to states of: Utah and Ohio; and OUS (International) to countries of: Italy, Netherlands, Argentina, Korea, Jordan, Spain, and Bangladesh.

Product description

KWIK-STIK and LYFO DISK, quality control kit for culture media Catalog # 0805K and 0805L

Code information (models, lots, serial numbers)

Lot #: 805-234-4, 805-234-5 UDI: (0805K) 30845357019531, (0805L) 10845357019544