Skip to content

Sprinter Cart/Sprinter Cart XL Infusion Pole. Component… recallZ-2193-2025 · Maquet Cardiopulmonary Gmbh · initiated Jun 27, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiopulmonary Gmbh or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device does not comply with the labeling requirements of international standard IEC 60601-1.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION NOTICE dated June 2025 was mailed to consignees. The notification informs consignees that Getinge will be sending a field service technician to visit and verify if a safety sign is present on affected devices. Consignees are asked to share the recall notification with users of affected devices or to facilities to which product was further distributed for their awareness. Consignees are to complete and return the provided response form to [email protected]. Consignees with any questions are to direct them to Technical Support at 1-877-943-8872 (options 4, 2, 1) or by email at [email protected].

Recall details

Recall number
Z-2193-2025
Status
Open, Classified
Initiated by the company
Jun 27, 2025
Posted by the FDA
Jul 29, 2025
Recalling firm
Maquet Cardiopulmonary Gmbh
Quantity
7,754 units
Product code
KRI
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, New Zealand, Canada, Hong Kong, India, Israel, Japan, Singapore, South Korea, Taiwan, Thailand, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Serbia, Slovakia, Slovenia, Spain, Sweden, Turkey, Reunion, Romania, Russian Federation, Switzerland, Ukraine, United Kingdom, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Mexico, Paraguay, Peru, Uruguay, Bahrain, Egypt, Iran, Iraq, Jordan, Kuwait, Lebanon, Oman, Pakistan, Qatar, Saudi Arabia, South Africa, United Arab Emirates, Yemen.

Product description

Sprinter Cart/Sprinter Cart XL Infusion Pole. Component used to hold 2 x 2 kg infusion bags.

Code information (models, lots, serial numbers)

Model No. 701033599; UDI: 4037691257860.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page