Skip to content

Bartels ELISA Legionella Urinary Antigen. Model Number:… recallZ-2292-2025 · Trinity Biotech USA · initiated Jul 22, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Trinity Biotech USA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 065 and 066 to contain particulate matter.

FDA root cause category: Under Investigation by firm

What the company is doing

Trinity Biotech initially notified consignees 06/05/2025 via letter. Consignees were instructed to inspect the conjugate bottles before use in both manual and automated testing and if any are observed with particulate matter, Trinity Biotech confirms the kit performs, and results are consistent with release specifications and IFU claims. However, replacement can be requested. Trinity Biotech notified consignees again on about 07/22/2025 via an updated URGENT: MEDICAL DEVICE Notification letter. Consignees were instructed that their trained laboratory technicians should inspect the conjugate bottles before use in both manual and automated testing, if any bottle is observed to contain conjugate with particulate matter, to discard the bottle and use a clear bottle available. If there are no other available clear bottles, please contact Trinity Biotech for a replacement kit. The current conjugate will be replaced with Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 067 once available.

Recall details

Recall number
Z-2292-2025
Status
Open, Classified
Initiated by the company
Jul 22, 2025
Posted by the FDA
Aug 12, 2025
Recalling firm
Trinity Biotech USA
Quantity
968 kits
Product code
MJH
FDA submissions
K991074
Distribution
Worldwide - US Nationwide distribution in the states of Maryland and Utah. The countries of France, Germany, Ireland, Israel, Kuwait, Latvia, UK.

Product description

Bartels ELISA Legionella Urinary Antigen. Model Number: B1029-440. In Vitro Diagnostic test kit for the presumptive diagnosis of past or current Legionnaires Disease.

Code information (models, lots, serial numbers)

Model Number: B1029-440. UDI-DI: 05391516744065. Lot Numbers: 065 and 066

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.