Medical device recalls: August 2025
The FDA posted 251 medical device recall records in August 2025, from 71 recall events by 60 companies. 38 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Boston Scientific Corporation · 4 events, 16 records, 15 Class I
- Baxter Healthcare Corporation · 3 events, 5 records, 4 Class I
- CareFusion 303, Inc. · 3 events, 4 records, 2 Class I
- MEDLINE INDUSTRIES, LP - Northfield · 3 events, 3 records, 1 Class I
- GE Medical Systems, LLC · 2 events, 24 records
- Beckman Coulter, Inc. · 2 events, 7 records
- Philips Ultrasound, Inc · 1 event, 69 records
- AVID Medical, Inc. · 1 event, 24 records
FDA root cause
- Device Design · 79
- Under Investigation by firm · 65
- Equipment maintenance · 25
- Process control · 23
- Nonconforming Material/Component · 11
- Labeling design · 8
- Process design · 6
- Labeling Change Control · 5
Class I recalls (38)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2332-2025 · Baxter Healthcare Corporation · posted Aug 29, 2025 · Open, Classified
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement…
- Z-2333-2025 · Baxter Healthcare Corporation · posted Aug 29, 2025 · Open, Classified
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement…
- Z-2334-2025 · Baxter Healthcare Corporation · posted Aug 29, 2025 · Open, Classified
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement…
- Z-2426-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2427-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2428-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2429-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2430-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2431-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2432-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2433-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2434-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2435-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2436-2025 · Boston Scientific Corporation · posted Aug 29, 2025 · Open, Classified
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring…
- Z-2460-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Aug 28, 2025 · Open, Classified
Specific lots of reprocessed electrophysiology catheters may contain small residual particulates.
- Z-2343-2025 · Tandem Diabetes Care, Inc. · posted Aug 27, 2025 · Open, Classified
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead…
- Z-2325-2025 · Boston Scientific Corporation · posted Aug 20, 2025 · Open, Classified
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
- Z-2326-2025 · Boston Scientific Corporation · posted Aug 20, 2025 · Open, Classified
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
- Z-2327-2025 · Boston Scientific Corporation · posted Aug 20, 2025 · Open, Classified
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
- Z-2328-2025 · Hamilton Medical AG · posted Aug 19, 2025 · Open, Classified
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an…
- Z-2173-2025 · Baxter Healthcare Corporation · posted Aug 14, 2025 · Open, Classified
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The…
- Z-2181-2025 · Ambu Inc. · posted Aug 13, 2025 · Open, Classified
Potential for the manometer port being blocked rendering the manometer non-functional.
- Brand Name: HeartMateClass IZ-2290-2025 · Thoratec LLC · posted Aug 13, 2025 · Open, Classified
Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
- Brand Name: HeartMateClass IZ-2291-2025 · Thoratec LLC · posted Aug 13, 2025 · Open, Classified
Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
- Z-2227-2025 · Philips Respironics, Inc. · posted Aug 8, 2025 · Open, Classified
Devices may possess a programming error resulting in an incorrect device configuration.
- Z-2228-2025 · Philips Respironics, Inc. · posted Aug 8, 2025 · Open, Classified
Devices may possess a programming error resulting in an incorrect device configuration.
- Z-2229-2025 · Philips Respironics, Inc. · posted Aug 8, 2025 · Open, Classified
Devices may possess a programming error resulting in an incorrect device configuration.
- Z-2273-2025 · Boston Scientific Corporation · posted Aug 8, 2025 · Open, Classified
Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device…
- Z-2129-2025 · ICU Medical, Inc. · posted Aug 6, 2025 · Open, Classified
ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
- Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.Class IZ-2176-2025 · B Braun Medical Inc · posted Aug 6, 2025 · Open, Classified
Label is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension set is not indicated for removal of air and does not feature an air vent, which is common to air eliminating…
- Z-2179-2025 · CareFusion 303, Inc. · posted Aug 6, 2025 · Open, Classified
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- BD Alaris Pump Infusion SetsClass IZ-2180-2025 · CareFusion 303, Inc. · posted Aug 6, 2025 · Open, Classified
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
- Z-2182-2025 · Flexicare Medical (Dongguan) Ltd. · posted Aug 1, 2025 · Open, Classified
Laryngoscope handles may not illuminate as intended.
- Z-2183-2025 · Flexicare Medical (Dongguan) Ltd. · posted Aug 1, 2025 · Open, Classified
Laryngoscope handles may not illuminate as intended.
- Z-2184-2025 · Flexicare Medical (Dongguan) Ltd. · posted Aug 1, 2025 · Open, Classified
Laryngoscope handles may not illuminate as intended.
- Z-2185-2025 · Flexicare Medical (Dongguan) Ltd. · posted Aug 1, 2025 · Open, Classified
Laryngoscope handles may not illuminate as intended.
- Z-2186-2025 · Flexicare Medical (Dongguan) Ltd. · posted Aug 1, 2025 · Open, Classified
Laryngoscope handles may not illuminate as intended.
- Z-2187-2025 · Flexicare Medical (Dongguan) Ltd. · posted Aug 1, 2025 · Open, Classified
Laryngoscope handles may not illuminate as intended.
Class II recalls (140)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2462-2025 · C.R. Bard Inc · posted Aug 28, 2025 · Open, Classified
Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney injury and/or hydronephrosis.
- Brand Name: HOMEDICSClass IIZ-2463-2025 · Zhejiang Luyao Electronics Technology Co., Ltd. · posted Aug 28, 2025 · Open, Classified
Regulatory documentation error and misleading color packaging claim.
- Z-2464-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2465-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2466-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2467-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2468-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2469-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2470-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2471-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2472-2025 · American Contract Systems Inc · posted Aug 28, 2025 · Open, Classified
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to…
- Z-2473-2025 · Diagnostica Stago, Inc. · posted Aug 28, 2025 · Open, Classified
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Z-2474-2025 · Diagnostica Stago, Inc. · posted Aug 28, 2025 · Open, Classified
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Z-2475-2025 · Diagnostica Stago, Inc. · posted Aug 28, 2025 · Open, Classified
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Z-2476-2025 · Diagnostica Stago, Inc. · posted Aug 28, 2025 · Open, Classified
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- BD PyxisClass IIZ-2477-2025 · CareFusion 303, Inc. · posted Aug 28, 2025 · Open, Classified
Antivirus software was not consistently installed on impacted devices during the implementation process.
- Z-2479-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Aug 28, 2025 · Open, Classified
Kits contain SureStep Foley Tray Systems recalled by BD that include incorrect product information insert sheets. If a mislabeled tray is used, patient risk ranges from minor injury to injury that requires medical intervention, such as infection or…
- Z-2452-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Aug 27, 2025 · Open, Classified
Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This…
- Z-2453-2025 · Bd Switzerland Sarl · posted Aug 27, 2025 · Open, Classified
Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it.
- IPC Powerease System REF 2300000Class IIZ-2461-2025 · Medtronic Xomed, Inc. · posted Aug 27, 2025 · Open, Classified
Due to out of the box wobble of the driver.
- Z-2437-2025 · Vantive US Healthcare LLC · posted Aug 26, 2025 · Open, Classified
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
- Z-2438-2025 · Vantive US Healthcare LLC · posted Aug 26, 2025 · Open, Classified
Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD…
- TherMax Blood Warmer Unit-US, PrisMax Accessory, Product Codes: 955630, 955515, 955702, 955631 ↗Class IIZ-2439-2025 · Vantive US Healthcare LLC · posted Aug 26, 2025 · Open, Classified
Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284: Thermax Disposable Not Inserted
- Z-2440-2025 · Orascoptic Surgical Acuity · posted Aug 26, 2025 · Open, Classified
Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.
- Z-2441-2025 · Orascoptic Surgical Acuity · posted Aug 26, 2025 · Open, Classified
Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.
- Z-2442-2025 · Orascoptic Surgical Acuity · posted Aug 26, 2025 · Open, Classified
Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.
- Z-2443-2025 · Orascoptic Surgical Acuity · posted Aug 26, 2025 · Open, Classified
Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.
- Z-2444-2025 · Orascoptic Surgical Acuity · posted Aug 26, 2025 · Open, Classified
Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.
- Z-2445-2025 · Orascoptic Surgical Acuity · posted Aug 26, 2025 · Open, Classified
Due to complaints, loupe nose pads containing nickel may cause an allergic reaction.
- Brand Name: Atlantis IO FLO-S ↗Class IIZ-2421-2025 · Dentsply IH, Inc. · posted Aug 25, 2025 · Open, Classified
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5…
- Brand Name: Atlantis IO FLO ↗Class IIZ-2422-2025 · Dentsply IH, Inc. · posted Aug 25, 2025 · Open, Classified
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5…
- ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS BOWED CANAL FILLING STEMS. Canal filling stem implants.Class IIZ-2423-2025 · Onkos Surgical, Inc. · posted Aug 25, 2025 · Open, Classified
Unidentified substance of unknown impact was identified on devices during a retrospective evaluation.
- Z-2424-2025 · Onkos Surgical, Inc. · posted Aug 25, 2025 · Open, Classified
Unidentified substance of unknown impact was identified on devices during a retrospective evaluation.
- Z-2425-2025 · Onkos Surgical, Inc. · posted Aug 25, 2025 · Open, Classified
Unidentified substance of unknown impact was identified on devices during a retrospective evaluation.
- GE HealthCare OEC 9800.Class IIZ-2341-2025 · GE Medical Systems, LLC · posted Aug 22, 2025 · Open, Classified
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small…
- Z-2342-2025 · GE Medical Systems, LLC · posted Aug 22, 2025 · Open, Classified
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small…
- Dynex Agility Sample Tips ↗Class IIZ-2345-2025 · Dynex Technologies, Inc. · posted Aug 22, 2025 · Open, Classified
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which…
- Z-2415-2025 · Baxter Healthcare Corporation · posted Aug 22, 2025 · Open, Classified
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied…
- Brand Name: THORClass IIZ-2416-2025 · THOR Photomedicine Ltd · posted Aug 22, 2025 · Open, Classified
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
- Brand Name: THORClass IIZ-2417-2025 · THOR Photomedicine Ltd · posted Aug 22, 2025 · Open, Classified
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
- Brand Name: THORClass IIZ-2418-2025 · THOR Photomedicine Ltd · posted Aug 22, 2025 · Open, Classified
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
- Brand Name: THOR ↗Class IIZ-2419-2025 · THOR Photomedicine Ltd · posted Aug 22, 2025 · Open, Classified
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
- Corin Operating TablesClass IIZ-2420-2025 · Maquet GMBH · posted Aug 22, 2025 · Open, Classified
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue…
- OtoLase Starter KitClass IIZ-2344-2025 · Lumenis, Ltd. · posted Aug 21, 2025 · Open, Classified
The potential for unsterilized product within finished product labeled as sterile.
- Z-2338-2025 · Olympus Corporation of the Americas · posted Aug 19, 2025 · Open, Classified
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
- Z-2339-2025 · Olympus Corporation of the Americas · posted Aug 19, 2025 · Open, Classified
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
- Z-2340-2025 · Boston Scientific Neuromodulation Corporation · posted Aug 19, 2025 · Open, Classified
Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may…
- Z-2336-2025 · Immunotech A.S. · posted Aug 18, 2025 · Open, Classified
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases…
- Clearview Strep A Exact II Dipstick Respiratory Kit, CLIA Waived. Catalog Number: 4581225020Class IIZ-2337-2025 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 18, 2025 · Open, Classified
Due to a classification error in the inventory management system, specific medical devices were shipped to customers that did not have the requisite license to purchase such devices.
- Pentax Medical Video ProcessorClass IIZ-2329-2025 · Pentax of America Inc · posted Aug 15, 2025 · Open, Classified
During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during…
- Brand Name: Q-StopClass IIZ-2330-2025 · Drs Vascular, Inc · posted Aug 15, 2025 · Open, Classified
XXX
- Z-2331-2025 · Spiggle & Theis Mt Gmbh · posted Aug 15, 2025 · Open, Classified
Affected product show elevated cytotoxicity values. Use of affected injection needles may present a risk of intolerance reaction or allergic reaction.
- Z-2335-2025 · Ortho-Clinical Diagnostics, Inc. · posted Aug 15, 2025 · Open, Classified
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially…
- Z-2298-2025 · Insightra Medical Inc · posted Aug 13, 2025 · Terminated
Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.
- Z-2299-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2300-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2301-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2302-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2303-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2304-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2305-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2306-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2307-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2308-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2309-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2310-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2311-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2312-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2313-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2314-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2315-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2316-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2317-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2318-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2319-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Z-2320-2025 · GE Medical Systems, LLC · posted Aug 13, 2025 · Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA…
- Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860Class IIZ-2322-2025 · Beckman Coulter, Inc. · posted Aug 13, 2025 · Open, Classified
Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients…
- HemosIL LMW Heparin ControlsClass IIZ-2323-2025 · Instrumentation Laboratory · posted Aug 13, 2025 · Open, Classified
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin…
- Z-2292-2025 · Trinity Biotech USA · posted Aug 12, 2025 · Open, Classified
Potential for Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 065 and 066 to contain particulate matter.
- Z-2293-2025 · Merit Medical Systems, Inc. · posted Aug 12, 2025 · Open, Classified
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
- Z-2294-2025 · Merit Medical Systems, Inc. · posted Aug 12, 2025 · Open, Classified
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
- Custom Manifold Kit:Class IIZ-2295-2025 · Merit Medical Systems, Inc. · posted Aug 12, 2025 · Open, Classified
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
- Custom Waste Management Kit:Class IIZ-2296-2025 · Merit Medical Systems, Inc. · posted Aug 12, 2025 · Open, Classified
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
- Z-2297-2025 · PIE Medical Imaging B.V. · posted Aug 12, 2025 · Open, Classified
When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to…
- Z-2287-2025 · Spark Biomedical Inc · posted Aug 11, 2025 · Open, Classified
Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.
- Z-2288-2025 · Waldemar Link GmbH & Co. KG (Mfg Site) · posted Aug 11, 2025 · Open, Classified
Inconsistent size terminology and color coding used on labeling
- Z-2289-2025 · Waldemar Link GmbH & Co. KG (Mfg Site) · posted Aug 11, 2025 · Open, Classified
Inconsistent size terminology and color coding used on labeling
- Z-2278-2025 · Boston Scientific Corporation · posted Aug 8, 2025 · Open, Classified
The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.
- Brand Name: PERLA TL MISClass IIZ-2279-2025 · Spineart SA · posted Aug 8, 2025 · Open, Classified
Cannulated fenestrated polyaxial screws may be incorrectly labeled
- Brand Name: PERLA TL MISClass IIZ-2280-2025 · Spineart SA · posted Aug 8, 2025 · Open, Classified
Cannulated fenestrated polyaxial screws may be incorrectly labeled
- Z-2281-2025 · Carefree Surgical Specialties · posted Aug 8, 2025 · Open, Classified
Due to two issues: 1. Product contamination (biological foreign matter) that could compromise sterility. 2. Incorrect quantity of gauze in sterile packaging
- Z-2282-2025 · Carefree Surgical Specialties · posted Aug 8, 2025 · Open, Classified
Due to two issues: 1. Product contamination (biological foreign matter) that could compromise sterility. 2. Incorrect quantity of gauze in sterile packaging
- Z-2283-2025 · Carefree Surgical Specialties · posted Aug 8, 2025 · Open, Classified
Due to two issues: 1. Product contamination (biological foreign matter) that could compromise sterility. 2. Incorrect quantity of gauze in sterile packaging
- Z-2284-2025 · Carefree Surgical Specialties · posted Aug 8, 2025 · Open, Classified
Due to two issues: 1. Product contamination (biological foreign matter) that could compromise sterility. 2. Incorrect quantity of gauze in sterile packaging
- Z-2285-2025 · Encore Medical, LP · posted Aug 8, 2025 · Open, Classified
Knee implants contain incorrect labeling (size and/or side incorrect)
- Z-2286-2025 · Encore Medical, LP · posted Aug 8, 2025 · Open, Classified
Knee implants contain incorrect labeling (size and/or side incorrect)
- Z-2277-2025 · Stryker Corporation · posted Aug 7, 2025 · Open, Classified
Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse…
- DASH SARS-CoV-2 & Flu A/B TestClass IIZ-2274-2025 · Nuclein LLC · posted Aug 6, 2025 · Open, Classified
Due to manufacturing error, assay test may result in missed diagnosis or delay of treatment. (Potential false negative result)
- Z-2275-2025 · Intersurgical Inc · posted Aug 6, 2025 · Open, Classified
Potential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue irritation, inflammation and infection.
- Z-2276-2025 · Intersurgical Inc · posted Aug 6, 2025 · Open, Classified
Potential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue irritation, inflammation and infection.
- Z-2266-2025 · Beckman Coulter, Inc. · posted Aug 5, 2025 · Open, Classified
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the…
- Z-2267-2025 · Beckman Coulter, Inc. · posted Aug 5, 2025 · Open, Classified
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the…
- Z-2268-2025 · Beckman Coulter, Inc. · posted Aug 5, 2025 · Open, Classified
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the…
- Z-2269-2025 · Beckman Coulter, Inc. · posted Aug 5, 2025 · Open, Classified
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the…
- Z-2270-2025 · Beckman Coulter, Inc. · posted Aug 5, 2025 · Open, Classified
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the…
- Z-2271-2025 · Beckman Coulter, Inc. · posted Aug 5, 2025 · Open, Classified
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the…
- Z-2272-2025 · Siemens Healthcare Diagnostics Inc · posted Aug 5, 2025 · Open, Classified
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14…
- Z-2240-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2241-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2242-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2243-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2244-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2245-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2246-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2247-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2248-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2249-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2250-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2251-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2252-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2253-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2254-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2255-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2256-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2257-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2258-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2259-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2260-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2261-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2262-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2263-2025 · AVID Medical, Inc. · posted Aug 4, 2025 · Open, Classified
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
- Z-2264-2025 · Beckman Coulter Inc. · posted Aug 4, 2025 · Open, Classified
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their…
- Z-2231-2025 · 3M Company · posted Aug 1, 2025 · Open, Classified
Notice was issued to clarify flow rates related to the inlet fluid temperature.
- Z-2232-2025 · Fresenius Kabi USA, LLC · posted Aug 1, 2025 · Open, Classified
Potential for external cassette leaks
- Z-2233-2025 · Fresenius Kabi USA, LLC · posted Aug 1, 2025 · Open, Classified
Potential for external cassette leaks
- LVP Primary Administration Set, Single Outlet Low Sorbing (Qty 25). Product Code: SET-0021-25.Class IIZ-2234-2025 · Fresenius Kabi USA, LLC · posted Aug 1, 2025 · Open, Classified
Potential for external cassette leaks
- Z-2235-2025 · Fresenius Kabi USA, LLC · posted Aug 1, 2025 · Open, Classified
Potential for external cassette leaks
- Z-2237-2025 · Becton, Dickinson and Company, BD Bio Sciences · posted Aug 1, 2025 · Open, Classified
Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
- Z-2238-2025 · Becton, Dickinson and Company, BD Bio Sciences · posted Aug 1, 2025 · Open, Classified
Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.
- Z-2239-2025 · Philips Medical Systems (Cleveland) Inc · posted Aug 1, 2025 · Open, Classified
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Class III recalls (73)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-2451-2025 · Bausch & Lomb Surgical, Inc. · posted Aug 26, 2025 · Open, Classified
An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected intraocular lens could result in functional visual impairment or reduction in visual acuity which could result in the need…
- 3D6-2 Transducer Probe ↗Class IIIZ-2346-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- 3D9-3v Transducer ProbeClass IIIZ-2347-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- 7.5MHZ Endo Transducer Probe ↗Class IIIZ-2348-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- BP10-5ec Transducer ProbeClass IIIZ-2349-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C10-3v Transducer ProbeClass IIIZ-2350-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C10-4ec Transducer ProbeClass IIIZ-2351-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- c5-1 Transducer ProbeClass IIIZ-2352-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C5-2 Transducer ProbeClass IIIZ-2353-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C5-2 Lumify Transducer ProbeClass IIIZ-2354-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C6-2 Transducer ProbeClass IIIZ-2355-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C6-3 Transducer ProbeClass IIIZ-2356-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C8-4v Transducer ProbeClass IIIZ-2357-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C8-4v Transducer ProbeClass IIIZ-2358-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C8-5 Transducer ProbeClass IIIZ-2359-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C9-2 Transducer ProbeClass IIIZ-2360-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C9-3io Transducer ProbeClass IIIZ-2361-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C9-3v Transducer ProbeClass IIIZ-2362-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C9-4 Transducer ProbeClass IIIZ-2363-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C9-4v Transducer ProbeClass IIIZ-2364-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- C9-5ec Transducer ProbeClass IIIZ-2365-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- D2cwc Transducer ProbeClass IIIZ-2366-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- D2TCD Transducer ProbeClass IIIZ-2367-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- D5cwc Transducer ProbeClass IIIZ-2368-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- eL18-4 Transducer ProbeClass IIIZ-2369-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L10-4LAP Transducer ProbeClass IIIZ-2370-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L11-3 Transducer ProbeClass IIIZ-2371-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L12-3 Transducer ProbeClass IIIZ-2372-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L12-4 Transducer ProbeClass IIIZ-2373-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L12-4 Lumify Transducer ProbeClass IIIZ-2374-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L12-5 38mm Transducer ProbeClass IIIZ-2375-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L12-5 50 Transducer ProbeClass IIIZ-2376-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L15-7io Transducer ProbeClass IIIZ-2377-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- Z-2378-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L17-5 Transducer ProbeClass IIIZ-2379-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L18-5 Transducer ProbeClass IIIZ-2380-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L8-4 Transducer ProbeClass IIIZ-2381-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- L9-3 Transducer ProbeClass IIIZ-2382-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- mC12-3 Transducer ProbeClass IIIZ-2383-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- mC7-2 Transducer ProbeClass IIIZ-2384-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- Mini Multi TEE Transducer ProbeClass IIIZ-2385-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- mL26-8 Transducer ProbeClass IIIZ-2386-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- OMNI II TEE Transducer Probe ↗Class IIIZ-2387-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- OMNI III TEE Transducer ProbeClass IIIZ-2388-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S12-4 Transducer ProbeClass IIIZ-2389-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S3-1 Transducer ProbeClass IIIZ-2390-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S4-1 Transducer ProbeClass IIIZ-2391-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S4-1 Lumify Transducer ProbeClass IIIZ-2392-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S4-2 Transducer ProbeClass IIIZ-2393-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S5-2 Transducer ProbeClass IIIZ-2394-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S7-3t Transducer ProbeClass IIIZ-2395-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S7-2t Transducer ProbeClass IIIZ-2396-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S8-3 Transducer ProbeClass IIIZ-2397-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S8-3t Transducer ProbeClass IIIZ-2398-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- S9-2 Transducer ProbeClass IIIZ-2399-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- SCNHD LA L7-4 HDI Transducer Probe ↗Class IIIZ-2400-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- Transducer C10-3v Transducer Probe ↗Class IIIZ-2401-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- TRANSDUCER L12-5 Transducer Probe ↗Class IIIZ-2402-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- V6-2 Transducer ProbeClass IIIZ-2403-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- V9-2 Transducer ProbeClass IIIZ-2404-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- VL13-5 Transducer ProbeClass IIIZ-2405-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X11-4t Transducer ProbeClass IIIZ-2406-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X3-1 Transducer Probe ↗Class IIIZ-2407-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X5-1 Transducer ProbeClass IIIZ-2408-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X5-1c Transducer ProbeClass IIIZ-2409-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X6-1 Transducer ProbeClass IIIZ-2410-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X7-2 Transducer ProbeClass IIIZ-2411-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X7-2t Transducer ProbeClass IIIZ-2412-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- X8-2t Transducer ProbeClass IIIZ-2413-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- XL14-3 Transducer ProbeClass IIIZ-2414-2025 · Philips Ultrasound, Inc · posted Aug 22, 2025 · Open, Classified
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
- Z-2324-2025 · Diversatek Healthcare · posted Aug 14, 2025 · Open, Classified
Anorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the expiration date. All inner labels on kit components display the correct expiration date.
- Description/REF:Class IIIZ-2321-2025 · CareFusion 303, Inc. · posted Aug 13, 2025 · Open, Classified
Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.
- Brand Name: T4 Total ELISAClass IIIZ-2265-2025 · DRG International, Inc. · posted Aug 4, 2025 · Open, Classified
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration…