i-view video laryngoscope. Model Number: 8008000 recallZ-2102-2025 · Intersurgical Inc · initiated Jun 17, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Intersurgical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for faulty devices as a result of depleted batteries.
FDA root cause category: Component design/selection
What the company is doing
Intersurgical notified consignees and account executives via email on 06/17/2025. The notification instructed consignees and account executives to immediately discontinue use and quarantine any affected stock on hand, notify customers if affected units have been further distributed, and contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization (RGA) Number and return label to return all affected units. Also, they were instructed to complete and return the Customer Product Recall Acknowledgement Form.
Recall details
- Recall number
- Z-2102-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 17, 2025
- Posted by the FDA
- Jul 11, 2025
- Recalling firm
- Intersurgical Inc
- Quantity
- 2,152 pieces
- Product code
- CCW
- Distribution
- US distribution AZ, CA, FL, GA, IL, IN, KS, MD, MI, MO, NV, NY, OH, PA, SC, TX, VA.
Product description
i-view video laryngoscope. Model Number: 8008000. The i-view video laryngoscope is indicated to facilitate laryngoscopy.
Code information (models, lots, serial numbers)
Model Number: 8008000. UDI-DI (Case): 05030267166497. UDI-DI (Single Unit): 5030267150649. Lot Numbers: 1240555, 1240793.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- I-VIEW VIDEO LARYNGOSCOPE 8008000Intersurgical Incorporated · Laryngoscope, Rigid1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.