Skip to content

Color Cuff recallZ-2277-2025 · Stryker Corporation · initiated Jul 2, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.

FDA root cause category: Equipment maintenance

What the company is doing

On July 2, 2025, the firm notified customers via Urgent Medical Device Recall letters. Customers were instructed to locate and quarantine any affected products at their facility and return an enclosed business reply form to the firm. Stryker will provide a shipping label to return the affected product, and credit will be issued to affected customers upon receipt of the recalled product.

Recall details

Recall number
Z-2277-2025
Status
Open, Classified
Initiated by the company
Jul 2, 2025
Posted by the FDA
Aug 7, 2025
Recalling firm
Stryker Corporation
Quantity
233.230
Product code
KCY
Distribution
Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom

Product description

Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235

Code information (models, lots, serial numbers)

5921-018-135 GTIN 07613154599257 5921-018-235 GTIN 07613154599295 5921-024-235 GTIN 07613154599370

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates