BD COR System Software. Model Number: 444829 recallZ-2162-2025 · Becton Dickinson & Co. · initiated Jun 23, 2025

Class II Open, Classified All recalls posted in July 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.

FDA root cause category: Under Investigation by firm

What the company is doing

Becton, Dickinson and Company began notification of consignees on about 06/23/2025 via telephone call and followed by a letter. Consignees were instructed to note the outlined workflows when loading specimens with the Over labeling feature enabled and consider retesting affected specimens for HPV testing if affected. Consignees were also instructed to post the notification within their facility, notify customers if affected units were further distributed, and complete and return the provided Customer Response Form.

Recall details

Recall number
Z-2162-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 23, 2025
Posted by the FDA
Jul 23, 2025
Recalling firm
Becton Dickinson & Co.
Quantity
7 systems
Product code
MAQ
FDA submissions
P160037
Distribution
Worldwide - US Nationwide distribution in the states of Florida, Texas and Michigan. The countries of Sweden and Belgium.

Product description

BD COR System Software. Model Number: 444829.

Code information (models, lots, serial numbers)

Model Number: 444829. UDI-DI: 00382904448295. Software version 8.80 and above.

Other devices under the same FDA submission

5 devices cleared under P160037 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.