Ambu SPUR II Labeled as the following: 1 SPUR II Adult… recallZ-2181-2025 · Ambu Inc. · initiated Jul 9, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ambu Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for the manometer port being blocked rendering the manometer non-functional.

FDA root cause category: Equipment maintenance

What the company is doing

Ambu notified consignees on about 07/09/2025 via Urgent Field Safety Notice letter. Consignees were instructed to identify if there are any affected units on hand, notify Ambu of how many affected units are in possession, discard those affected units, and complete and return Appendix 1 of the Field Safety Notice. Additionally, consignees were instructed to notify all applicable personnel and notify customers if further distributed.

Recall details

Recall number
Z-2181-2025
Status
Open, Classified
Initiated by the company
Jul 9, 2025
Posted by the FDA
Aug 13, 2025
Recalling firm
Ambu Inc.
Quantity
87,156 units
Product code
BTM
FDA submissions
K152931
Distribution
Worldwide distribution - US Nationwide and the country of Canada.

Product description

Ambu SPUR II Labeled as the following: 1 SPUR II Adult Resuscitator, Adult, Catalog Number: 523211000 2 SPUR II Adult Resuscitator w/Mercury Filter, Adult, Catalog Number: 523611051 3 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 523611057 4 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 524611000 5 SPUR II Adult Resuscitator, Pop-off open 40 ,w/PEEP Valve 20, Adult, Catalog Number: 524611001 6 SPUR II Adult Resuscitator w/PEEP Valve 20, Adult, Catalog Number: 524611011 7 SPUR II Adult Resuscitator w/Pressure Limiting Valve, Adult, Catalog Number: 524611031 8 SPUR II Adult Resuscitator w/Pressure Limiting Valve, Adult, Catalog Number: 524611047 9 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 524611051 10 SPUR II Pediatric Resuscitator w/10 O2 tube & Manometer, Catalog Number: 530200016 11 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530212000 12 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530212001 13 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213000 14 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213001 15 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213011 16 SPUR II Pediatric Resuscitator w/Manometer, Pediatric, Catalog Number: 530213030 17 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213031 18 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530213048 19 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530214000 20 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530613000 21 SPUR II Pediatric Resuscitator w/Manometer & PEEP Valve 20, Pediatric, Catalog Number: 530613031 22 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530613071 23 SPUR II Pediatric Resuscitator w/PEEP Valve 20,Manometer & CO2 Detector, Pediatric, Catalog Number: 530613831 24 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530614017 25 SPUR II Pediatric Resuscitator w/Manometer, Pediatric, Catalog Number: 530614030 26 SPUR II Pediatric Resuscitator w/CO2 Detector, Pediatric, Catalog Number: 530614800 27 SPUR II Pediatric Resuscitator, INF& TOD W/Manometer, Pediatric, Catalog Number: 530615030 28 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 530619031 29 SPUR II Pediatric Resuscitator w/Expiratory Filter, Manometer PEEP Valve, Pediatric, Catalog Number: 531600051 30 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613000 31 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613001 32 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613010 33 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613031 34 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531613047 35 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number:531613051 36 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531614026 37 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number: 531638000 38 SPUR II Adult Resuscitator, Single Patient Use Resuscitator, Adult, Catalog Number: 523611051E 39 SPUR II Pediatric Resuscitator, Single Patient Use Resuscitator, Pediatric, Catalog Number:530213000B The Ambu SPUR II Resuscitator is a single patient use resuscitator intended for pulmonary resus

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

1. Catalog Number: 523211000; UDI Number: 05707480152193; Lot Numbers: 1001113554, 1001118764. 2. Catalog Number: 523611051; UDI Number: 05707480149513; Lot Numbers: 2000015874. 3. Catalog Number: 523611057; UDI Number: 05707480038688; Lot Numbers: 2000015876. 4. Catalog Number: 524611000; UDI Number: 05707480152278; Lot Numbers: 1001110297. 5. Catalog Number: 524611001; UDI Number: 05707480149698; Lot Numbers: 2000015881. 6. Catalog Number: 524611011; UDI Number: 05707480149711; Lot Numbers: 2000015882. 7. Catalog Number: 524611031; UDI Number: 05707480149735; Lot Numbers: 2000015884, 2000015968. 8. Catalog Number: 524611047; UDI Number: 05707480149759; Lot Numbers: 2000016102. 9. Catalog Number: 524611051; UDI Number: 05707480154395; Lot Numbers: 2000015969, 2000016252. 10. Catalog Number: 530200016; UDI Number: 05707480149773; Lot Numbers: 2000014933, 2000015857. 11. Catalog Number: 530212000; UDI Number: 05707480152353; Lot Numbers: 1001113557. 12. Catalog Number: 530212001; UDI Number: 05707480149797; Lot Numbers: 2000016408. 13. Catalog Number: 530213000; UDI Number: 05707480152377; Lot Numbers: 1001110299, 1001113558, 1001118767, 1001106634. 14. Catalog Number: 530213001; UDI Number: 05707480152391; Lot Numbers: 1001106635. 15. Catalog Number: 530213011; UDI Number: 05707480152438; Lot Numbers: 1001118768. 16. Catalog Number: 530213030; UDI Number: 05707480149858; Lot Numbers: 2000016409. 17. Catalog Number: 530213031; UDI Number: 05707480152452; Lot Numbers: 1001113560, 1001118769. 18. Catalog Number: 530213048; UDI Number: 05707480154159; Lot Numbers: 2000016529, 2000016410. 19. Catalog Number: 530214000; UDI Number: 05707480152490; Lot Numbers: 1001106636, 1001113561. 20. Catalog Number: 530613000; UDI Number: 05707480152612; Lot Numbers: 1001110301, 1001121455, 1001106637, 1001118771. 21. Catalog Number: 530613031; UDI Number: 05707480150212; Lot Numbers: 2000015889. 22. Catalog Number: 530613071; UDI Number: 05707480150298; Lot Numbers: 2000015574. 23. Catalog Number: 530613831; UDI Number: 05707480150373; Lot Numbers: 2000015179, 2000015726, 2000015891. 24. Catalog Number: 530614017; UDI Number: 05707480150397; Lot Numbers: 2000015440. 25. Catalog Number: 530614030; UDI Number: 05707480150410; Lot Numbers: 2000014938, 2000015299. 26. Catalog Number: 530614800; UDI Number: 05707480150434; Lot Numbers: 2000015098. 27. Catalog Number: 530615030; UDI Number: 05707480150472; Lot Numbers: 2000015893. 28. Catalog Number: 530619031; UDI Number: 05707480150595; Lot Numbers: 2000016761. 29. Catalog Number: 531600051; UDI Number: 05707480163441; Lot Numbers: 2000015181, 2000015577. 30. Catalog Number: 531613000; UDI Number: 05707480152773; Lot Numbers: 1001118773, 1001121460, 1001106640. 31. Catalog Number: 531613001; UDI Number: 05707480152797; Lot Numbers: 1001106641, 1001110305, 1001113567, 1001118774. 32. Catalog Number: 531613010; UDI Number: 05707480152810; Lot Numbers: 1001113568. 33. Catalog Number: 531613031; UDI Number: 05707480150717; Lot Numbers: 2000015182, 2000015731, 2000015896, 2000016112, 2000016263. 34. Catalog Number: 531613047; UDI Number: 05707480150755; Lot Numbers: 2000015444, 2000015580, 2000015897, 2000015982. 35. Catalog Number: 531613051; UDI Number: 05707480154470; Lot Numbers: 2000016825. 36. Catalog Number: 531614026; UDI Number: 05707480150854; Lot Numbers: 2000016265, 2000016612. 37. Catalog Number: 531638000; UDI Number: 05707480150878; Lot Numbers: 2000015184, 2000015583. 38. Catalog Number: 523611051E; UDI Number: 05707480154135; Lot Numbers: 2000015292, 2000015559, 2000016401. 39. Catalog Number: 530213000B; UDI Number: 05707480153534; Lot Numbers: 1001113559

Devices named in this recall

38 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page