Skip to content

3M Ranger recallZ-2231-2025 · 3M Company · initiated Apr 22, 2025

Class II Open, Classified All recalls posted in August 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from 3M Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Notice was issued to clarify flow rates related to the inlet fluid temperature.

FDA root cause category: Device Design

What the company is doing

Solventum issued an Urgent Medical Device Correction notice dated 04/21/2025 to its consignees via email and US certified mail. The notice explained the issue, the modification related to the inlet fluid temperature to address the issue, and forthcoming instructions for use updates. All users of the impacted products are being asked to take the following actions. " Ensure this notification is passed on to all those who need to be aware within your organization and to any department where the affected products have been transferred. " Continue to use the products as indicated, considering the updated information provided. " Keep this notice on hand until you are notified that the updated IFUs and operator s manuals are available. " Complete the attached Customer Acknowledgement Form acknowledging that you have read and understood this letter and the updated information provided. " Email the completed form to [email protected]. For additional questions or need to report an adverse event, contact the Solventum [formerly 3M Health Care] Customer Helpline at 1-800-228-3957, option 6.

Recall details

Recall number
Z-2231-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 22, 2025
Posted by the FDA
Aug 1, 2025
Recalling firm
3M Company
Quantity
186,880 units
Product code
LGZ
FDA submissions
K060939
Distribution
US (nationwide) and Argentina, Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, Canary Islands, China, Costa Rica, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Korea, Luxembourg, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.

Product description

3M Ranger Irrigation Fluid Warming Set, REF 24750

Code information (models, lots, serial numbers)

UDI/DI 50707387792720 (box), 10707387792722 (pouch, )SKU 7100235070, All lots with a manufacturing date after March 2022

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates