Sparrow Ascent Patient Controller, REF's: 11-110-K… recallZ-2287-2025 · Spark Biomedical Inc · initiated Jul 2, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spark Biomedical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.

FDA root cause category: Process control

What the company is doing

On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions: 1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working. 2. Current devices may continue to be used until they are replaced. 3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives. 4. Please notify all appropriate personnel at your facility that need to know about this recall.

Recall details

Recall number
Z-2287-2025
Status
Open, Classified
Initiated by the company
Jul 2, 2025
Posted by the FDA
Aug 11, 2025
Recalling firm
Spark Biomedical Inc
Quantity
78 units
Product code
PZR
FDA submissions
K230796
Distribution
US Nationwide distribution.

Product description

Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28

Code information (models, lots, serial numbers)

Lot numbers: V2408xxxx and runs through V2519xxxx UDI: 11-110/00850052017033; 11-310-K/00860005396948; 11-320-K/00860005396986; 11-321-K/00850052017163; 11-310-K-28/00850052017088; 11-320-K-28/00850052017071; 11-321-K-28/00850052017170