epoc BGEM BUN Test Card [25pk]. Material Number: 10736515 recallZ-2272-2025 · Siemens Healthcare Diagnostics Inc · initiated Jul 7, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthcare Diagnostics notified consignees via letter on about 07/07/2025. Consignees were instructed to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, discard and request a replacement of any affected lots or if your institution does not need to report sodium, you can turn off sodium following the instruction of the owner s manual. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form included with letter, notify customers if products were further distributed, and to retain this letter with your laboratory records and forward this letter to those who may have received this product at your site.
Recall details
- Recall number
- Z-2272-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 7, 2025
- Posted by the FDA
- Aug 5, 2025
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 37,554 units
- Product code
- JGS
- FDA submissions
- K200107
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.
Product description
epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Material Number: 10736515. GTIN: 00809708121860. Lot Numbers: 02-25095-30, 02-25096-30, 02-25098-30, 02-25099-30, 02-25100-30, 02-25101-30, 02-25102-30, 02-25103-30, 02-25105-30, 02-25111-30, 02-25112-30, 02-25114-30, 02-25115-30, 02-25116-30, 02-25117-30, 02-25119-30, 02-25121-30, 02-25122-30, 02-25124-30, 02-25096-40, 02-25097-40, 02-25098-40, 02-25099-40, 02-25101-40, 02-25102-40, 02-25105-40, 02-25106-40, 02-25121-40, 02-25122-40, 02-25124-40, 02-25125-40, 02-25096-50, 02-25098-50, 02-25099-50, 02-25100-50, 02-25102-50, 02-25103-50, 02-25105-50, 02-25107-50, 02-25108-50, 02-25111-50, 02-25112-50, 02-25113-50, 02-25114-50, 02-25115-50, 02-25119-50, 02-25120-50, 12-25096-60, 12-25097-60, 12-25098-60, 12-25100-60, 12-25101-60, 12-25102-60, 12-25103-60, 12-25111-60, 02-25112-60, 02-25113-60, 02-25114-60, 02-25115-60, 02-25116-60, 02-25117-60, 02-25118-60, 02-25120-60, 02-25121-60, 02-25122-60, 02-25123-60, 02-25124-60, 02-25099-70, 02-25101-70, 02-25104-70, 02-25105-70, 02-25107-70, 02-25108-70, 02-25110-70, 02-25111-70, 02-25112-70, 02-25113-70, 02-25115-70, 02-25116-70, 02-25118-70, 02-25120-70.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EPOC BGEM BUN TEST CARD 10736515Epocal Inc6 recalls1 report
Other devices under the same FDA submission
1 device cleared under K200107 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- epoc NXS Host 11413475Epocal Inc · ACID, LACTIC, ENZYMATIC METHOD2 recalls6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.