Brand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS recallZ-2286-2025 · Encore Medical, LP · initiated Jul 9, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Knee implants contain incorrect labeling (size and/or side incorrect)
FDA root cause category: Labeling Change Control
What the company is doing
On January 15, 2026 Enovis (Encore Medical, L.P.) issued a Urgent: Medical Device Recall expansion notification to affected consignees via Email. The expansion notification included additional lots of affected product. Initial notification was sent out on July 9, 2025. Enovis asked consignees to take the following actions: 1. Review your stock for the products and lot numbers for the items to be returned. 2. Remove any indicated product from stock and quarantine. 3. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 4. Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through your regional email address, to place a replacement request. You will be provided with an RMA number and shipping label to return the affected product. 5. Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service.
Recall details
- Recall number
- Z-2286-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 9, 2025
- Posted by the FDA
- Aug 8, 2025
- Recalling firm
- Encore Medical, LP
- Quantity
- 40 implants
- Product code
- JWH
- FDA submissions
- K143242
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IA, IL, IN, KS, LA, NY, TN, TX.
Product description
Brand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEE INS, 7R 16MM, VE Model/Catalog Number: 342-16-707 Product Description: Enovis knee devices are intended for treatment of patients who are candidates for knee arthroplasty per the Indication for use. While total knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
Code information (models, lots, serial numbers)
Lot Code: Item: 342-16-707 GTIN: 00888912167642 Lot: 161T1053, 161T1054
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DJO SURGICAL 342-16-707Encore Medical, L.P. · Guide, Surgical, Instrument1 recall9 reports
Other devices under the same FDA submission
25 devices cleared under K143242 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- DJO SURGICAL 342-11-706Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER8 reports
- DJO SURGICAL 342-13-709Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER7 reports
- DJO SURGICAL 341-11-708Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER5 reports
- DJO SURGICAL 341-11-709Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER5 reports
- DJO SURGICAL 342-11-708Encore Medical, L.P. · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive4 reports
- DJO SURGICAL 341-11-704Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 reports
- DJO SURGICAL 342-13-705Encore Medical, L.P. · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive3 reports
- DJO SURGICAL 341-13-706Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 reports
- DJO SURGICAL 342-11-707Encore Medical, L.P. · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive3 reports
- DJO SURGICAL 341-13-704Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- DJO SURGICAL 341-11-706Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- DJO SURGICAL 342-13-707Encore Medical, L.P. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97188).
- Z-2285-2025 · Open, ClassifiedBrand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEE INS, 4L 16MM, VE Model/Catalog Number: 341-16-704 Product Description: Enovis knee devices are inte…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page