DAVINCI UROLOGY. Medical convenience kit recallZ-2245-2025 · AVID Medical, Inc. · initiated Jun 11, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.
FDA root cause category: Under Investigation by firm
What the company is doing
Consignees were sent an URGENT: MEDICAL DEVICE FIELD CORRECTION notice, dated 6/11/25, via email. The notification instructs consignees to not use the CLEARIFY Visualization System (component #5303907) and discard. All other components in the kit may be continued to be used. Consignees are to add warning labels to affected procedure trays to ensure the affected component is discarded. The provided Recall Response Form is to be completed and returned to Owens&Minor. Any quality problems or adverse events associated with recalled devices may be reported via email to [email protected]
Recall details
- Recall number
- Z-2245-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 11, 2025
- Posted by the FDA
- Aug 4, 2025
- Recalling firm
- AVID Medical, Inc.
- Quantity
- 166 units
- Product code
- LRO
- Distribution
- Domestic: CA, IL, MD, OH, PA, TX, WA.
Product description
DAVINCI UROLOGY. Medical convenience kit.
Code information (models, lots, serial numbers)
Model No. WRNM038-09, WRNM038-10, WRNM038-11; UDI: 10809160369302, 10809160418338, 10809160432259; Kit Lot No. 1589886, 1593087, 1606686, 1606688, 1615262, 1611133, 1627007.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Halyard WRNM038-09_0009Avid Medical, Inc. · General surgery tray1 recall0 reports
- Halyard WRNM038-10_0002Avid Medical, Inc. · General surgery tray1 recall0 reports
- Halyard WRNM038-11_0003Avid Medical, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97149).
- Z-2240-2025 · Open, ClassifiedLAPAROSCOPIC TRAY. Medical convenience kit.
- Z-2241-2025 · Open, ClassifiedCARDIAC ROBOT PACK. Medical convenience kit.
- Z-2242-2025 · Open, ClassifiedCT DAVINCI. Medical convenience kit.
- Z-2243-2025 · Open, ClassifiedDA VINCI PACK URO/PROSTATE. Medical convenience kit.
- Z-2244-2025 · Open, ClassifiedDAVINCI TRUCUSTOM BUNDLE. Medical convenience kit.
- Z-2246-2025 · Open, ClassifiedGEN LAPAROSCOPY PACK. Medical convenience kit.
- Z-2247-2025 · Open, ClassifiedGENERAL LAPAROSCOPY TRAY. Medical convenience kit.
- Z-2248-2025 · Open, ClassifiedGENERAL ROBOTIC PACK. Medical convenience kit.
- Z-2249-2025 · Open, ClassifiedGS ENDOSCOPY PACK. Medical convenience kit.
- Z-2250-2025 · Open, ClassifiedGYN LAPAROSCOPY PACK. Medical convenience kit.
- Z-2251-2025 · Open, ClassifiedGYN ROBOTIC PACK. Medical convenience kit.
- Z-2252-2025 · Open, ClassifiedHVI CARDIAC ROBOT PACK. Medical convenience kit.
- Z-2253-2025 · Open, ClassifiedLAP CHOLE PACK. Medical convenience kit.
- Z-2254-2025 · Open, ClassifiedLAPAROSCOPIC GYN. Medical convenience kit.
- Z-2255-2025 · Open, ClassifiedLAPAROSCOPIC PROSTATECTOMY. Medical convenience kit.
- Z-2256-2025 · Open, ClassifiedLAPAROSCOPY PACK. Medical convenience kit.
- Z-2257-2025 · Open, ClassifiedLAVH PACK. Medical convenience kit.
- Z-2258-2025 · Open, ClassifiedMAJOR THORACOSCOPY BASIN. Medical convenience kit.
- Z-2259-2025 · Open, ClassifiedPACK GENERAL ROBOTIC. Medical convenience kit.
- Z-2260-2025 · Open, ClassifiedPARTIAL NEPHRECTOMY - ROBOTIC. Medical convenience kit.
- Z-2261-2025 · Open, ClassifiedPARTIAL NEPHRECTOMY PACK. Medical convenience kit.
- Z-2262-2025 · Open, ClassifiedROBOTIC PROSTATECTOMY PACK. Medical convenience kit.
- Z-2263-2025 · Open, ClassifiedROBOTIC URO/GYN PACK. Medical convenience kit.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page