Skip to content

FARAWAVE recallZ-2278-2025 · Boston Scientific Corporation · initiated Jun 25, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An Urgent Medical Device Removal - Immediate Action Required notification letter dated June 2025 was sent to customers. Actions 1. Do NOT use affected product (Attachment 1), remove affected devices from your facility s inventory, segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page five. 5. Return affected product. Email: [email protected] Reach out to your local Boston Scientific representative with any questions.

Recall details

Recall number
Z-2278-2025
Status
Open, Classified
Initiated by the company
Jun 25, 2025
Posted by the FDA
Aug 8, 2025
Recalling firm
Boston Scientific Corporation
Quantity
445 units
Product code
QZI
FDA submissions
P230030
Distribution
US Nationwide.

Product description

FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401

Code information (models, lots, serial numbers)

UDI-DI: 00191506043148; Batch Numbers: 36543581 36543644 36543646 36568553 36572565 36572568 36572571 36572576 36578308 36578314 36579110 36598352 36598809 36599402 36600406 36600543 36600544 36600545 36600578

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

12 devices cleared under P230030 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.