Breathing circuit set, coaxial recallZ-2167-2025 · Hamilton Medical AG · initiated Jun 26, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

FDA root cause category: Process control

What the company is doing

Starting on 6/26/2025, recall notices were emailed to customers who were asked to do the following: Return affected devices. If one of the affected breathing circuits would be already in use with a patient, it is requested to exchange it and discard the product after replacement. If you are a distributor, please forward to any organization where potentially affected product has been transferred. Complete and return the acknowledgement form via email to [email protected] If you have any questions contact firm at: [email protected] and 800-426-6331

Recall details

Recall number
Z-2167-2025
Status
Open, Classified
Initiated by the company
Jun 26, 2025
Posted by the FDA
Jul 30, 2025
Recalling firm
Hamilton Medical AG
Quantity
2,560
Product code
BZO
Distribution
US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

Product description

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Code information (models, lots, serial numbers)

UDI-DI: 07630002802963, Lots: 199675, 199676