PrisMax V2-US, Hemodialysis Delivery System, Product Codes:… recallZ-2437-2025 · Vantive US Healthcare LLC · initiated Jul 28, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Vantive US Healthcare LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
FDA root cause category: Under Investigation by firm
What the company is doing
Vantive issued an Urgent Medical Device Correction notice to it consignees on 0728/2025 via USPS first class mail. The notice explained the issues with the devices, hazard involved, and requested the following: Actions to be Taken by Customers 1. Operators may continue to safely use the PrisMax Systems and TherMax Blood Warmer Units per the associated Operator s Manuals and on-screen instructions. If further assistance is required, please contact your local Vantive account representative. Please ensure that all operators of these devices are made aware of this Urgent Medical Device Correction notification. Vantive recommends that a copy of this notification is posted within the appropriate location of your facility. 2. As corrections become available, a local Vantive representative will contact your facility to determine the correction plan and schedule the associated correction(s) for impacted devices. 3. If you received this communication directly from Vantive, complete the enclosed Vantive Customer Reply Form and return it to Vantive by scanning and e-mailing it to [email protected]. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive an email and/or phone call from Vantive to confirm your receipt of this notification. 4. If you purchased this product from a distributor, please note that responding using the Vantive Customer Reply Form is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your
Recall details
- Recall number
- Z-2437-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 28, 2025
- Posted by the FDA
- Aug 26, 2025
- Recalling firm
- Vantive US Healthcare LLC
- Quantity
- 4772 units
- Product code
- KDI
- FDA submissions
- K190910
- Distribution
- Worldwide
Product description
PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701
Code information (models, lots, serial numbers)
UDI 00085412639499, All Serial Numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- prisMaxPrisMax System 955626Baxter Healthcare Corporation · Dialyzer, high permeability with or without sealed dialysate system4 recalls37 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.