DASH SARS-CoV-2 & Flu A/B Test recallZ-2274-2025 · Nuclein LLC · initiated Jun 25, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nuclein LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing error, assay test may result in missed diagnosis or delay of treatment. (Potential false negative result)

FDA root cause category: Process control

What the company is doing

On June 25, 2025, Nuclein, LLC issued a Urgent: Medical Device Recall notification to affected customers. On July 8, 2025 an updated notification was issued via E.Mail. a. Review your inventory of DASH SARS-CoV-2 & Flu A/B Tests to confirm the quantity of Lot BB02 or BB03 that you have remaining. b. Suspend the use and quarantine the affected lot of DASH SARS-CoV-2 & Flu A/B Tests; Nuclein will replace any unused affected product. c. Review all results of the product lots detailed in Section 6, below, to identify potentially erroneous diagnosis, except in those cases where alternative confirmation was obtained. d. Forward this information to all individuals and departments within your organization using the product information detailed in Section 6. If you are not the end user, forward this notice to the product end user. e. Complete the Acknowledgement and Receipt Form below and either email the completed form to [email protected] or mail to the address identified in the Acknowledgement and Receipt Form. f. Return the affected tests that you have remaining; Nuclein Customer Support will follow up with you on the details of returning or destroying the affected product.

Recall details

Recall number
Z-2274-2025
Status
Open, Classified
Initiated by the company
Jun 25, 2025
Posted by the FDA
Aug 6, 2025
Recalling firm
Nuclein LLC
Quantity
710 units
Product code
QOF
FDA submissions
K241652
Distribution
US distribution to: Florida, New York, Texas, and West Virginia

Product description

DASH SARS-CoV-2 & Flu A/B Test Model/Catalog Number: SG-0006 combination COVID-19 and Flu test. Each DASH SARS-CoV-2 & Flu A/B Test is individually pouched and labeled. The tests are provided to users in quantities of 10 per kit. Used on the DASH instrument.

Code information (models, lots, serial numbers)

Lot Code: Lot: BB02 UDI: (01)00850063063036(17)260220(10)BB02 Expiration Date: 20 Feb. 2026 Lot Code: Lot: BB03 UDI: (01)00850063063036(17)260220(10)BB03 Expiration Date: 20 Feb. 2026

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K241652 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.