BostonSight PROSE Lens recallZ-2191-2025 · Boston Foundation For Sight · initiated Jun 10, 2025
Class II Completed All recalls posted in July 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Foundation For Sight or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Select orders of lenses were manufactured with errors impacting sagittal height, haptics, or both.
FDA root cause category: Process control
What the company is doing
Consignees were notified via email with the subject "URGENT: BostonSight SCLERAL- Recall of Selected Orders". Consignees are to return affected devices to BostonSight via the prepaid envelopes that are to be provided. Replacement units will be provided, and the firm will be in contact with information regarding replacements. Consignees with any questions are to contact Laurel Villers by email at [email protected] or by phone at 781-726-7436.
Recall details
- Recall number
- Z-2191-2025
- Recall class
- Class II
- Status
- Completed
- Initiated by the company
- Jun 10, 2025
- Posted by the FDA
- Jul 29, 2025
- Recalling firm
- Boston Foundation For Sight
- Quantity
- 38 units
- Product code
- HQD
- FDA submissions
- K161461
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada.
Product description
BostonSight PROSE Lens.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1. UDI Number [Lot Number], Serial Numbers: (01)00785811435077(10)24341007 [24341007] Serial Numbers: 534720, 536887, 540283. 2. UDI Number [Lot Number], Serial Numbers: (01)00785811435639(10)20030007 [20030007] Serial Number: 534761. 3. UDI Number [Lot Number], Serial Numbers: (01)00785811435639(10)24214004 [24214004] Serial Numbers: 536327, 537399, 538805. 4. UDI Number [Lot Number], Serial Numbers: (01)00785811435677(10)22078002 [22078002] Serial Numbers: 535453. 5. UDI Number [Lot Number], Serial Numbers: (01)00785811435677(10)24183007 [24183007] Serial Numbers: 540811. 6. UDI Number [Lot Number], Serial Numbers: (01)00865576000201(10)F2402477 [F2402477] Serial Numbers: 532389, 532390, 532660, 532666, 532816, 533417, 533418, 533419, 533420. 7. UDI Number [Lot Number], Serial Numbers: (01)00865576000201(10)F2403286 [F2403286] Serial Numbers: 534129, 534130, 535028, 535029, 535711, 535712, 535869, 536897, 537916, 537918, 538203. 8. UDI Number [Lot Number], Serial Numbers: (01)00865576000201(10)F2403377 [F2403377] Serial Numbers: 533357, 533358. 9. UDI Number [Lot Number], Serial Numbers: (01)00865576000201(10)T020007 [)T020007] Serial Numbers: 539865. 10. UDI Number [Lot Number], Serial Numbers: (01)00865576000218(10)F2300341 [F2300341] Serial Numbers: 535908. 11. UDI Number [Lot Number], Serial Numbers: (01)00865576000294(10)F2402750 [F2402750] Serial Numbers: 532817, 534107, 536034. 12. UDI Number [Lot Number], Serial Numbers: (01)00865576000294(10)F2403018 [F2403018] Serial Numbers: 540171. 13. UDI Number [Lot Number], Serial Numbers: (01)00865576000294(10)F2500408 [F2500408] Serial Numbers: 539392.
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Boston XO2 hexafocon BBausch & Lomb Incorporated · Lens, Contact (Other Material) - Daily3 recalls1 report
- Optimum Extra roflufocon DContamac Holdings Limited · Lens, Contact (Other Material) - Daily3 recalls1 report
- Optimum Extreme roflufocon EContamac Holdings Limited · Lens, Contact (Other Material) - Daily3 recalls0 reports
- Boston Equalens II oprifocon ABausch & Lomb Incorporated · Lens,Contact(Rigid Gas Permeable)-Extended Wear2 recalls1 report
- Optimum Infinite Tisilfocon AContamac Holdings Limited · Lens, Contact (Other Material) - Daily2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97123).
- Z-2190-2025 · CompletedBostonSight SCLERAL Lens
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page