Skip to content

Brand Name: Ultra IABP Catheter Kit recallZ-2298-2025 · Insightra Medical Inc · initiated Jul 19, 2025

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Insightra Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.

FDA root cause category: No Marketing Application

What the company is doing

On June 19, 2025, Insightra Medical issued a recall notification to affected consignees via postal mail and email. On August 4, 2025, Insightra Medical issued a "Urgent Medical Device Recall" notification update via E-Mail to provide additional information concerning the labeling. Insightra ask consignees to take the following actions: 1. Inform consignees to discard any previously printed versions of the IFU. 2.Review your firm s procedures and processes to ensure that they are up to date with the updated IFU. 3.If your firm has further distributed the devices to other facilities, please provide that facility with a copy of the notification. 4.Ensure that all clinicians and users are informed of the IFU updates. 5.Access the latest IFU via the QR code located inside the product box lid or via web address: https://tn-mi.com/wp-content/uploads/2023/06/D02-0043-Ultra-IAB-IFU-2019.pdf 6.Retain the letter with your product documentation. 7.Please email the acknowledgement form to us at [email protected]. 8. Report any device concerns to Insightra. Please note that adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Recall details

Recall number
Z-2298-2025
Status
Terminated
Initiated by the company
Jul 19, 2025
Posted by the FDA
Aug 13, 2025
Recalling firm
Insightra Medical Inc
Quantity
24227 units
Product code
DSP
FDA submissions
K082746
Distribution
Worldwide - US Nationwide distribution in the states of TX, MO, FL, MI and the countries of Albania, Azerbaijan, Bangladesh, Brazil, Cambodia, Colombia, Hong Kong, India, Indonesia, Iran, Israel, Saudi Arabia, Kuwait, Lebanon, Malaysia, Nepal, Oman, Pakistan, Pakistan, Peru, Romania, Russia, Serbia, Singapore, United Arab Emirates.

Product description

Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Description: The Insightra Ultra IABP Catheter Kit is a 7 French catheter available with either a 40, 35, 30, 25, or 20 cc balloon. The catheter has two lumens and is designed to provide counterpulsation cardiac assist therapy. The outer lumen is a channel for helium used to inflate and deflate the balloon. The inner lumen is used for a guidewire and blood pressure measurement. It is packaged as a convenience kit with accessories (one-way valve, 50 cc syringe, guide wires, introducer needle, dilator, pressure tubing, Datascope & Arrow adaptors). The balloon inflation and deflation cycle is synchronized (using a commercially available control console) with the ECG or arterial pressure to provide counterpulsation synchronized with the heartbeat. It is intended to increase coronary perfusion, decrease the workload of the left ventricle and allow healing of the myocardium Component: No

Code information (models, lots, serial numbers)

Lot code: All lots / UDI: IMU7F-40: 00850012832546; IMU7F-35: 00850012832515; IMU7F-30: 00850012832485; IMU7F-25: None; IMU7F-20: None

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates