Skip to content

PERLA TL MIS recallZ-2279-2025 · Spineart SA · initiated Jun 30, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spineart SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cannulated fenestrated polyaxial screws may be incorrectly labeled

FDA root cause category: Labeling Change Control

What the company is doing

On July 4, 2025, Spineart SA, issued a Urgent: Medical Device Recall Notification to affected consignees via E-Mail. On July 10, 2025, Spineart SA issued a revised notification to provide additional information. On July 16,2025, Spineart expanded the recall to include an additional product and lot. Spineart SA ask consignees to take the following steps: (1) Please acknowledge receipt of this letter by emailing to [email protected] (2) Please discontinue use and return any product on hand that is listed in the Product and Distribution Information section of this letter to: Spineart U.S.A. 23332 Mill Creek Drive Suite 150 Laguna Hills, CA. 92653 (3) You may have further distributed this product; please identify concerned customers and notify them at once of this product recall by using this document. (4) Surgeons/physicians should closely monitor patients that have been implanted with these potential non-conforming screws.

Recall details

Recall number
Z-2279-2025
Status
Open, Classified
Initiated by the company
Jun 30, 2025
Posted by the FDA
Aug 8, 2025
Recalling firm
Spineart SA
Quantity
16 units
Product code
NKB
FDA submissions
K203506
Distribution
US Nationwide distribution in the states of Florida , Kentucky, California.

Product description

Brand Name: PERLA TL MIS Product Name: CANNULATED FENESTRATED POLYAXIAL SCREW Model/Catalog Number: MPF-PS 65 40-S Software Version: Not Applicable Product Description: CANNULATED FENESTRATED POLYAXIAL SCREW Component: No

Code information (models, lots, serial numbers)

Lot Code: Lot number 8-4528 GTIN 07640305160561

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97248).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page