Skip to content

Access Thyroglobulin assay, a simultaneous one-step… recallZ-2322-2025 · Beckman Coulter, Inc. · initiated Jul 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification dated 7/2/25 was sent to customers. ACTION: " Discontinue using Access Thyroglobulin reagent lot 439163 and discard all remaining reagent packs from this lot. " At the discretion of the medical director, a retrospective review of patient results generated with Access Thyroglobulin reagent lot 439163 should be performed, which includes reviewing patient history, to assess the need for retesting. " Please contact your local Beckman Coulter representative for replacement product requests and to receive updates on availability and delivery. RESOLUTION: Beckman Coulter is investigating the root cause of this issue to prevent similar occurrences. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: " From our website: http://www.beckmancoulter.com " Contact your local Beckman Coulter Representative for replacement.

Recall details

Recall number
Z-2322-2025
Status
Open, Classified
Initiated by the company
Jul 10, 2025
Posted by the FDA
Aug 13, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
453 units
Product code
MSW
FDA submissions
K031269
Distribution
Worldwide - US Nationwide and the countries of Brazil, Canada, Georgia, Italy, Malaysia, Mexico, Philippines, Spain.

Product description

Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860

Code information (models, lots, serial numbers)

UDI-DI: 15099590227173; Lot Number: 439163

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates