Skip to content

PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H)… recallZ-2555-2025 · Orthofix U.S. LLC · initiated Jun 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthofix U.S. LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label side flap and inner label side flap information box. The 37 mm width on the front labels is correct.

FDA root cause category: Labeling design

What the company is doing

On 6/24/2025, recall notes were emailed to customers who were asked to do the following: 1) Examine your inventory and quarantine any affected product. 2) If implanted, healthcare Providers should determine whether it is necessary to inform the patient, based on clinical judgment and individual circumstances. 3) This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred. 4) Complete and return the response form to the firm. Should you have any questions regarding this matter, please contact the firm at [email protected]

Recall details

Recall number
Z-2555-2025
Status
Open, Classified
Initiated by the company
Jun 24, 2025
Posted by the FDA
Sep 11, 2025
Recalling firm
Orthofix U.S. LLC
Quantity
3
Product code
MAX
FDA submissions
K240749
Distribution
US Nationwide distribution in the states of CO, CA.

Product description

PILLAR SA Ti, STERILE, 37mm (W) x 28mm (D) x 18mm (H), 20-degrees, REF: 82-3713SP

Code information (models, lots, serial numbers)

UDI-DI: 18257200161010, Lots: 001, 002, and 003

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K240749 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page