Medical device recalls: September 2025
The FDA posted 189 medical device recall records in September 2025, from 82 recall events by 72 companies. 44 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Medical Systems Nederland B.V. · 4 events, 17 records
- Boston Scientific Corporation · 3 events, 7 records
- Siemens Medical Solutions USA, Inc · 2 events, 16 records, 15 Class I
- Medline Industries, LP · 2 events, 7 records, 7 Class I
- Abbott Point Of Care Inc. · 2 events, 4 records
- Maquet Cardiovascular, LLC · 2 events, 4 records
- Boston Scientific Neuromodulation Corporation · 2 events, 2 records
- Philips North America Llc · 1 event, 10 records
FDA root cause
- Under Investigation by firm · 75
- Device Design · 31
- Process control · 21
- Software design · 13
- No Marketing Application · 7
- Component design/selection · 7
- Incorrect or no expiration date · 6
- Component change control · 5
Class I recalls (44)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2654-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2655-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2656-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2657-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2658-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2659-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2660-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2661-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2662-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2663-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2664-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2665-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2666-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2667-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2668-2025 · Siemens Medical Solutions USA, Inc · posted Sep 30, 2025 · Open, Classified
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium…
- Z-2592-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Convenience kits labeled as sterile have not gone through the sterilization process.
- Z-2593-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Convenience kits labeled as sterile have not gone through the sterilization process.
- Z-2610-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or…
- Z-2611-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or…
- Z-2612-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or…
- Z-2613-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or…
- Z-2614-2025 · Medline Industries, LP · posted Sep 26, 2025 · Open, Classified
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or…
- Z-2566-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Sep 19, 2025 · Open, Classified
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
- Z-2567-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Sep 19, 2025 · Open, Classified
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.
- Z-2568-2025 · Abiomed, Inc. · posted Sep 19, 2025 · Open, Classified
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
- Z-2483-2025 · 3M Company · posted Sep 18, 2025 · Open, Classified
Field action to clarify the flow rates and inlet fluid temperatures, along with location of the outlet temperature for Ranger Blood/Fluid Warming High Flow Sets.
- Z-2538-2025 · Max Mobility LLC · posted Sep 12, 2025 · Open, Classified
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
- Z-2539-2025 · Max Mobility LLC · posted Sep 12, 2025 · Open, Classified
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
- Z-2540-2025 · Max Mobility LLC · posted Sep 12, 2025 · Open, Classified
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
- Z-2484-2025 · Olympus Corporation of the Americas · posted Sep 5, 2025 · Open, Classified
Potential for undetected, deformed a-traumatic tips.
- Z-2446-2025 · Dexcom, Inc. · posted Sep 4, 2025 · Open, Classified
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier),…
- Z-2447-2025 · Dexcom, Inc. · posted Sep 4, 2025 · Open, Classified
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier),…
- Z-2448-2025 · Dexcom, Inc. · posted Sep 4, 2025 · Open, Classified
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier),…
- Z-2449-2025 · Dexcom, Inc. · posted Sep 4, 2025 · Open, Classified
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier),…
- Z-2450-2025 · Dexcom, Inc. · posted Sep 4, 2025 · Open, Classified
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier),…
- Z-2480-2025 · Medtronic Perfusion Systems · posted Sep 3, 2025 · Open, Classified
The catheters may not retain their shape.
- Z-2481-2025 · Medtronic Perfusion Systems · posted Sep 3, 2025 · Open, Classified
The catheters may not retain their shape.
- Z-2482-2025 · Medtronic Perfusion Systems · posted Sep 3, 2025 · Open, Classified
The catheters may not retain their shape.
- Z-2454-2025 · mo-Vis BVBA · posted Sep 2, 2025 · Open, Classified
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
- Z-2455-2025 · mo-Vis BVBA · posted Sep 2, 2025 · Open, Classified
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
- Z-2456-2025 · mo-Vis BVBA · posted Sep 2, 2025 · Open, Classified
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
- Z-2457-2025 · mo-Vis BVBA · posted Sep 2, 2025 · Open, Classified
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
- Z-2458-2025 · mo-Vis BVBA · posted Sep 2, 2025 · Open, Classified
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
- Z-2459-2025 · mo-Vis BVBA · posted Sep 2, 2025 · Open, Classified
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Class II recalls (144)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2673-2025 · MedicalCommunications GmbH · posted Sep 30, 2025 · Open, Classified
Potential that the measured value may be smaller than the actual area.
- Z-2671-2025 · bioMerieux, Inc. · posted Sep 29, 2025 · Open, Classified
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
- Z-2672-2025 · Schiller, Ag · posted Sep 29, 2025 · Open, Classified
Vital sign monitoring instrument may trigger an error message during blood pressures measurements, Specifically, when the initial inflation pressure is set in the high-pressure range.
- Z-2637-2025 · North American Rescue LLC. · posted Sep 26, 2025 · Open, Classified
Device used for unilateral or bilateral occlusion of blood flow to the lower limbs has a potential manufacturing defect that may compromise the structural integrity of the device and device performance during use.
- Z-2638-2025 · Bard Peripheral Vascular Inc · posted Sep 26, 2025 · Open, Classified
Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is…
- Z-2639-2025 · Zap Surgical Systems · posted Sep 26, 2025 · Open, Classified
If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the…
- Z-2640-2025 · Howmedica Osteonics Corp. · posted Sep 26, 2025 · Open, Classified
a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
- Z-2622-2025 · Mindray DS USA, Inc. dba Mindray North America · posted Sep 25, 2025 · Open, Classified
Potential for activation of an abnormal alarm pause.
- Allura Xper FD10Class IIZ-2623-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD10/10Class IIZ-2624-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD10 OR TableClass IIZ-2625-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20Class IIZ-2626-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20 BiplaneClass IIZ-2627-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20 OR TableClass IIZ-2628-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Z-2629-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20/20Class IIZ-2630-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20/20 OR TableClass IIZ-2631-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20/10Class IIZ-2632-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20/15Class IIZ-2633-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura Xper FD20/15 OR TableClass IIZ-2634-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Allura CV20 ↗Class IIZ-2635-2025 · Philips Medical Systems Nederland B.V. · posted Sep 25, 2025 · Open, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual…
- Z-2636-2025 · Maquet Cardiovascular, LLC · posted Sep 25, 2025 · Open, Classified
Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and…
- Z-2620-2025 · Stryker Communications · posted Sep 24, 2025 · Open, Classified
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
- Z-2621-2025 · Tandem Diabetes Care, Inc. · posted Sep 24, 2025 · Open, Classified
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
- Z-2618-2025 · Boston Scientific Corporation · posted Sep 23, 2025 · Open, Classified
The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
- Z-2619-2025 · Biomet, Inc. · posted Sep 23, 2025 · Open, Classified
The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.
- Ortoma Inserter Adapter OTD:Class IIZ-2615-2025 · Ortoma AB · posted Sep 22, 2025 · Open, Classified
Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.
- Z-2594-2025 · Baxter Healthcare Corporation · posted Sep 19, 2025 · Open, Classified
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
- Z-2595-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2596-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2597-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2598-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2599-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2600-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2601-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2602-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2603-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2604-2025 · Philips North America Llc · posted Sep 19, 2025 · Open, Classified
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
- Z-2605-2025 · Ethicon, LLC · posted Sep 19, 2025 · Open, Classified
Potential for barb non-engagement.
- Z-2606-2025 · CooperVision, Inc. · posted Sep 19, 2025 · Open, Classified
One lot manufactured with an invalid sterilization cycle.
- Z-2607-2025 · CooperVision, Inc. · posted Sep 19, 2025 · Open, Classified
One lot manufactured with an invalid sterilization cycle.
- Voyant 1-Day Premium Toric, (stenfilcon A) contact lens. Model Number: Voyant 1-Day Premium Toric. ↗Class IIZ-2608-2025 · CooperVision, Inc. · posted Sep 19, 2025 · Open, Classified
One lot manufactured with an invalid sterilization cycle.
- Z-2609-2025 · Biofire Defense · posted Sep 19, 2025 · Open, Classified
FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.
- Z-2571-2025 · BigTree Sales Inc. · posted Sep 18, 2025 · Open, Classified
The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the…
- Z-2586-2025 · Maquet Cardiovascular, LLC · posted Sep 18, 2025 · Open, Classified
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- Z-2587-2025 · Maquet Cardiovascular, LLC · posted Sep 18, 2025 · Open, Classified
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- Z-2588-2025 · Maquet Cardiovascular, LLC · posted Sep 18, 2025 · Open, Classified
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
- Z-2589-2025 · C.R. Bard Inc · posted Sep 18, 2025 · Open, Classified
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection…
- Z-2590-2025 · C.R. Bard Inc · posted Sep 18, 2025 · Open, Classified
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection…
- Z-2591-2025 · C.R. Bard Inc · posted Sep 18, 2025 · Open, Classified
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection…
- Z-2572-2025 · Beckman Coulter Inc. · posted Sep 17, 2025 · Open, Classified
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- Z-2573-2025 · Beckman Coulter Inc. · posted Sep 17, 2025 · Open, Classified
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- Z-2574-2025 · Beckman Coulter Inc. · posted Sep 17, 2025 · Open, Classified
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- Z-2575-2025 · Beckman Coulter Inc. · posted Sep 17, 2025 · Open, Classified
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- Z-2576-2025 · Folsom Metal Products, Inc. · posted Sep 17, 2025 · Open, Classified
Labeling includes shelf life that has not been validated.
- Z-2577-2025 · Folsom Metal Products, Inc. · posted Sep 17, 2025 · Open, Classified
Labeling includes shelf life that has not been validated.
- Z-2578-2025 · Folsom Metal Products, Inc. · posted Sep 17, 2025 · Open, Classified
Labeling includes shelf life that has not been validated.
- Z-2579-2025 · Folsom Metal Products, Inc. · posted Sep 17, 2025 · Open, Classified
Labeling includes shelf life that has not been validated.
- Z-2580-2025 · Folsom Metal Products, Inc. · posted Sep 17, 2025 · Open, Classified
Labeling includes shelf life that has not been validated.
- Frontier Devices, REF: 301.916S1, 16 mm Distraction Pin, 10 single packs , Rx Only, Sterile ↗Class IIZ-2581-2025 · Folsom Metal Products, Inc. · posted Sep 17, 2025 · Open, Classified
Labeling includes shelf life that has not been validated.
- Z-2582-2025 · Abbott Point Of Care Inc. · posted Sep 17, 2025 · Open, Classified
Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.
- Z-2583-2025 · Abbott Point Of Care Inc. · posted Sep 17, 2025 · Open, Classified
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
- Z-2584-2025 · Abbott Point Of Care Inc. · posted Sep 17, 2025 · Open, Classified
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
- Z-2585-2025 · Abbott Point Of Care Inc. · posted Sep 17, 2025 · Open, Classified
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
- Z-2569-2025 · Medical Information Technology, Inc. · posted Sep 16, 2025 · Open, Classified
Entering multiple keys that trigger input simultaneously may remove data from first field of screens and/or questionnaires with more than one field.
- Z-2570-2025 · Philips Medical Systems (Cleveland) Inc · posted Sep 16, 2025 · Open, Classified
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
- Z-2563-2025 · Agilent Technologies Denmark ApS · posted Sep 15, 2025 · Open, Classified
Their is the potential for weak staining which may result in false negative CD20 identification.
- Z-2564-2025 · Agilent Technologies Denmark ApS · posted Sep 15, 2025 · Open, Classified
Their is the potential for weak staining which may result in false negative CD20 identification.
- Z-2565-2025 · Agilent Technologies Denmark ApS · posted Sep 15, 2025 · Open, Classified
Their is the potential for weak staining which may result in false negative CD20 identification.
- Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09Class IIZ-2558-2025 · Boston Scientific Neuromodulation Corporation · posted Sep 12, 2025 · Open, Classified
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience…
- Z-2559-2025 · Ortho-Clinical Diagnostics, Inc. · posted Sep 12, 2025 · Open, Classified
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay…
- Z-2560-2025 · Ortho-Clinical Diagnostics, Inc. · posted Sep 12, 2025 · Open, Classified
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay…
- Z-2561-2025 · Exactech, Inc. · posted Sep 12, 2025 · Open, Classified
Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.
- Z-2562-2025 · Exactech, Inc. · posted Sep 12, 2025 · Open, Classified
Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.
- Z-2478-2025 · Wasatch Photonics · posted Sep 11, 2025 · Open, Classified
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could…
- Z-2550-2025 · Boston Scientific Corporation · posted Sep 11, 2025 · Open, Classified
Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.
- Z-2551-2025 · Boston Scientific Corporation · posted Sep 11, 2025 · Open, Classified
Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.
- Brand Name: ReQuest Measles IgMClass IIZ-2552-2025 · Quest International, Inc. · posted Sep 11, 2025 · Open, Classified
Measles IgM Test Kit lacks premarket approval or clearance.
- 10Fore Hemostasis ValveClass IIZ-2553-2025 · Merit Medical Systems, Inc. · posted Sep 11, 2025 · Open, Classified
Due to a manufacturing issue involving the silicone quad ring where the ring is deformed or otherwise damaged and may cause a foreign body in the fluid pathway.
- Z-2554-2025 · Seaspine Orthopedics Corporation · posted Sep 11, 2025 · Open, Classified
Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use…
- Z-2555-2025 · Orthofix U.S. LLC · posted Sep 11, 2025 · Open, Classified
A 37 mm implant width intervertebral body fusion device has a labeling discrepancy with incorrect width of 33 mm stated on the outer label side flap and inner label side flap information box. The 37 mm width on the front labels is correct.
- Z-2556-2025 · Omnia Medical · posted Sep 11, 2025 · Open, Classified
Failure of fusion system instruments in the field.
- Z-2557-2025 · Omnia Medical · posted Sep 11, 2025 · Open, Classified
Failure of fusion system instruments in the field.
- Z-2545-2025 · Quidel Corporation · posted Sep 10, 2025 · Open, Classified
Dipstick strep A test has potential for false positive results.
- Brand Name: DJO SURGICALClass IIZ-2546-2025 · Encore Medical, LP · posted Sep 10, 2025 · Open, Classified
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
- Brand Name: DJO SURGICALClass IIZ-2547-2025 · Encore Medical, LP · posted Sep 10, 2025 · Open, Classified
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
- Brand Name: DJO SURGICALClass IIZ-2548-2025 · Encore Medical, LP · posted Sep 10, 2025 · Open, Classified
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
- Brand Name: DJO SURGICALClass IIZ-2549-2025 · Encore Medical, LP · posted Sep 10, 2025 · Open, Classified
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
- Z-2543-2025 · Philips Medical Systems Nederland B.V. · posted Sep 9, 2025 · Open, Classified
Software issue that results in the display of outdated information.
- Z-2544-2025 · Noxbox Ltd · posted Sep 9, 2025 · Open, Classified
Replacement of the internal Sample Pump due to high failure rates.
- Z-2541-2025 · Waldemar Link GmbH & Co. KG (Mfg Site) · posted Sep 8, 2025 · Open, Classified
Some modular LinkSymphoKnee TiNbN Tibial Components were manufactured with two uncoated security screws, which is incorrect.
- Z-2542-2025 · Siemens Healthcare Diagnostics, Inc. · posted Sep 8, 2025 · Open, Classified
Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.
- Brand Name: CELSITEClass IIZ-2528-2025 · B.Braun Medical S.A.S. · posted Sep 5, 2025 · Open, Classified
The potential for minor injury such as damage to the vessel and/or tearing of the vessel wall.
- Z-2529-2025 · Siemens Medical Solutions USA, Inc · posted Sep 5, 2025 · Open, Classified
Limited system movements after startup .
- Z-2530-2025 · AVID Medical, Inc. · posted Sep 5, 2025 · Open, Classified
Potential for open header bag seals, compromising sterility.
- Z-2531-2025 · AVID Medical, Inc. · posted Sep 5, 2025 · Open, Classified
Potential for open header bag seals, compromising sterility.
- Z-2532-2025 · Contract Medical International GmbH · posted Sep 5, 2025 · Open, Classified
Units may have been packaged with an incorrect pouch label, which lists the reference number and unique device identifier of the 4F version of the device, rather than the packaged 6F version.
- Z-2533-2025 · Boston Scientific Corporation · posted Sep 5, 2025 · Open, Classified
The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected…
- Z-2534-2025 · Boston Scientific Corporation · posted Sep 5, 2025 · Open, Classified
The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected…
- Z-2535-2025 · Boston Scientific Corporation · posted Sep 5, 2025 · Open, Classified
The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected…
- Z-2536-2025 · Boston Scientific Corporation · posted Sep 5, 2025 · Open, Classified
The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected…
- Z-2537-2025 · Corneat Vision, Ltd. · posted Sep 5, 2025 · Open, Classified
Retroactive; Due to complaints, labeling correction for permanent tissue-integrating surgical matrices to include prevention and management of would dehiscence.
- Z-2515-2025 · Bio-Rad Laboratories, Inc. · posted Sep 4, 2025 · Open, Classified
Due to a risk of false positive results that could lead to unnecessary medical treatment.
- Z-2516-2025 · Premia Spine Ltd · posted Sep 4, 2025 · Open, Classified
Potential for missing pins at tip of inserter.
- Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 ↗Class IIZ-2518-2025 · Merge Healthcare, Inc. · posted Sep 4, 2025 · Open, Classified
Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during…
- Z-2519-2025 · Lumicell, Inc. · posted Sep 4, 2025 · Open, Classified
The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.
- Z-2520-2025 · O&M Halyard Inc · posted Sep 4, 2025 · Open, Classified
Surgical drape packs may have open seals, which may compromise the sterility of the product.
- MI Cancer Seek REF MSC000Class IIZ-2521-2025 · Caris Life Sciences · posted Sep 4, 2025 · Open, Classified
Due to an incorrect test results provided that indicated the incorrect drug therapy recommendation
- Z-2522-2025 · ICU Medical Inc. · posted Sep 4, 2025 · Open, Classified
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Z-2523-2025 · ICU Medical Inc. · posted Sep 4, 2025 · Open, Classified
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Z-2524-2025 · ICU Medical Inc. · posted Sep 4, 2025 · Open, Classified
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Z-2525-2025 · ICU Medical Inc. · posted Sep 4, 2025 · Open, Classified
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Z-2526-2025 · ICU Medical Inc. · posted Sep 4, 2025 · Open, Classified
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Z-2527-2025 · ICU Medical Inc. · posted Sep 4, 2025 · Open, Classified
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- Z-2502-2025 · Philips Medical Systems Nederland B.V. · posted Sep 3, 2025 · Open, Classified
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
- Z-2503-2025 · Philips Medical Systems Nederland B.V. · posted Sep 3, 2025 · Open, Classified
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
- Z-2504-2025 · Philips Medical Systems Nederland B.V. · posted Sep 3, 2025 · Open, Classified
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
- Brand Name: MediHoneyClass IIZ-2505-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: MediHoneyClass IIZ-2506-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: MediHoneyClass IIZ-2507-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: MediHoneyClass IIZ-2508-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: MediHoneyClass IIZ-2509-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: MediHoneyClass IIZ-2510-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: MediHoneyClass IIZ-2511-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Brand Name: CVS Honey PadClass IIZ-2512-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Sep 3, 2025 · Open, Classified
Packaging failures were identified which could lead to a breach in the sterile barrier.
- Z-2513-2025 · Medshape, INC. · posted Sep 3, 2025 · Open, Classified
Bone external fixation system may have the pin that can become dislodged resulting in frame instability, the use of this product may cause a delay in surgery or a revision if the device is use.
- Z-2514-2025 · Preat Corp · posted Sep 3, 2025 · Open, Classified
Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation,…
- Superion IDS KitClass IIZ-2485-2025 · Boston Scientific Neuromodulation Corporation · posted Sep 2, 2025 · Open, Classified
As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.
- RefleXion X1 Model RXM1000Class IIZ-2486-2025 · Reflexion Medical, Inc. · posted Sep 2, 2025 · Open, Classified
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
- Z-2487-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2488-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2489-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2490-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2491-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2492-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2493-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2494-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2495-2025 · Raysearch Laboratories AB · posted Sep 2, 2025 · Open, Classified
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-2496-2025 · Medtronic MiniMed, Inc. · posted Sep 2, 2025 · Open, Classified
Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery,…
- Z-2497-2025 · Physio-Control, Inc. · posted Sep 2, 2025 · Open, Classified
Due to required inspections not being performed on products/units that have gone through servicing.
- Z-2498-2025 · Physio-Control, Inc. · posted Sep 2, 2025 · Open, Classified
Due to required inspections not being performed on products/units that have gone through servicing.
- Z-2499-2025 · Physio-Control, Inc. · posted Sep 2, 2025 · Open, Classified
Due to required inspections not being performed on products/units that have gone through servicing.
- Z-2500-2025 · Visgeneer, Inc. · posted Sep 2, 2025 · Open, Classified
Due to products distributed without premarket clearance or approvals.
- Z-2501-2025 · Visgeneer, Inc. · posted Sep 2, 2025 · Open, Classified
Due to products distributed without premarket clearance or approvals.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- FloPatch FP120, REF: FP120-FOT01-005Class IIIZ-2517-2025 · Flosonics Medical (R/A 1929803 ONTARIO CORP.) · posted Sep 4, 2025 · Open, Classified
Flowmeter patch as a human readable expiry date on the outer shipping packaging that reads 2025-06-20 when it should read 2026-06-20.