OtoLase Starter Kit recallZ-2344-2025 · Lumenis, Ltd. · initiated Jul 22, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Lumenis, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for unsterilized product within finished product labeled as sterile.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On July 22, 2025, URGENT MEDICAL DEVICE REMOVAL - IMMEDIATE ACTION REQUIRED letters were sent to customers. Instructions: " Immediately discontinue use and segregate the affected products to prevent accidental use and follow provided removal instructions on the attachment. " Post this information in a visible location near where the affected products are stored to ensure this information is readily accessible to all handlers and users of the device. Share this notice with any healthcare professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. " If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. Any adverse events or quality concerns associated with use of these devices should be reported to Boston Scientific via email at [email protected] or FDA s MedWatch Adverse Event Reporting program [www.fda.gov/MedWatch/report.htm or 1.800.FDA.1088 (332.1088)]. We appreciate your understanding as we take action to address this issue. At Boston Scientific, patient safety and customer satisfaction are our priority. We are committed to continuing to offer products that meet the quality standards that you expect from Boston Scientific. Contact your local Boston Scientific representative with any questions.
Recall details
- Recall number
- Z-2344-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 22, 2025
- Posted by the FDA
- Aug 21, 2025
- Recalling firm
- Lumenis, Ltd.
- Quantity
- 13 units
- Product code
- GEX
- FDA submissions
- K130164
- Distribution
- Domestic: PA, WI; International: Germany, India;
Product description
OtoLase Starter Kit; UPN: AC-2003748. Indicated for a variety of surgical uses including ablation, coagulation, incision, excision, and vaporization of soft tissue.
Code information (models, lots, serial numbers)
UPN: AC-2003748; UDI-DI: 07290109142807; Batch Numbers: 32644549 and 32769567; Expiration Date: 22Aug2025;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- OtoLase™ Starter KitLumenis Ltd. · Powered Laser Surgical Instrument1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page