Skip to content

FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a… recallZ-2609-2025 · Biofire Defense · initiated Jul 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biofire Defense or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.

FDA root cause category: Component design/selection

What the company is doing

On 7/10/2025, correction notices were mailed and emailed to customers who were asked to do the following: - If testing with only whole blood and/or sputum, there is no impact to the product s performance. Continue using panel kits as planned. - If testing with positive blood culture may occur, discontinue use, dispose of impacted kit(s) according to your lab s procedures, and only use replacement kits provided, or use impacted lots only with whole blood and/or sputum sample types. - Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties who may use this product, including others to whom you may have transferred our product. - Complete and return the acknowledgment of receipt form via email to [email protected]

Recall details

Recall number
Z-2609-2025
Status
Open, Classified
Initiated by the company
Jul 10, 2025
Posted by the FDA
Sep 19, 2025
Recalling firm
Biofire Defense
Quantity
130 Kits
Product code
PRD
FDA submissions
K170883
Distribution
US Nationwide distribution in the states of AL, AZ, UT, CA, TX, DC , VA, FL, MD, OH, HI, NY, KY, LA, MA, MI, MN, MO, NC, NH, RI, NV, TN, IA, WA.

Product description

FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system.

Code information (models, lots, serial numbers)

UDI-DI: 00851458005136. Lot(Expiration): 324324D(2026-02-19), D241022(2026-01-22), 319524D(2025-11-20), 315424D(2025-10-09), 315224D(2025-10-03), 315324D(2025-10-08)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page