Skip to content

Brand Name: THOR recallZ-2416-2025 · THOR Photomedicine Ltd · initiated Jul 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from THOR Photomedicine Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.

FDA root cause category: Process control

What the company is doing

On July 24, 2025, IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions by THOR: A service agent representing THOR will contact you within the next 10 weeks to schedule the fitting of the four high tensile strength steel bolts to the inside of the chassis. The process is likely to take around 3 hours and, of course, there is no fee. Required Actions for NovoTHOR Owners: 1. Please reply to this email within four days to acknowledge your receipt and understanding of this notification. 2. If you no longer own the NovoTHOR, please forward this notice to the new owner if the device was sold and reply to this email with the contact details of the new owner. Other Actions for Owners: Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm should the failure occur whilst they are lifting or lowering the canopy. " Ensure you warn all users of the potential risk " Ensure someone is present to lift and lower the canopy " Ensure the canopy is lifted and lowered gently and slowly " Ensure someone is present in the room or immediately outside the room during use. THOR Contact Information: Call +44 (0)1494 797 100, Monday through Friday, 5:00 AM to 2:00 PM, Eastern Time or via thorlaser.com/contact.htm.

Recall details

Recall number
Z-2416-2025
Status
Open, Classified
Initiated by the company
Jul 24, 2025
Posted by the FDA
Aug 22, 2025
Recalling firm
THOR Photomedicine Ltd
Quantity
80 units (66 US, 14 OUS)
Product code
ILY
Distribution
Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.

Product description

Brand Name: THOR Product Name: NovoTHOR Model/Catalog Number: S2188 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: No

Code information (models, lots, serial numbers)

Model/Catalog Number: S2188; UDI-DI: 05060494130503; Serial Numbers: 4004, 4015, 4017, 4019, 4021, 4022, 4024, 4026, 4028, 4029, 4035, 4036, 4037, 4042, 4043, 4044, 4046, 4047, 4049, 4054, 4055, 4059, 4066, 4071, 4072, 4073, 4074, 4076, 4077, 4078, 4079, 4082, 4083, 4084, 4085, 4086, 4087, 4088, 4090, 4091, 4094, 4095, 4097, 4099, 4101, 4105, 4106, 4111, 4112, 4114, 4116, 4117, 4121, 4123, 4124, 4125, 4126, 4130, 4140, 4141, 4143, 4144, 4151, 4153, 4157, 4158, 4009, 4020, 4103, 4107, 4109, 4110, 4113, 4120, 4122, 4127, 4133, 4134, 4138, 4139;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97297).

Sources and updates