ViziShot 2FLEX. Single Use Aspiration Needle. Model Number:… recallZ-2484-2025 · Olympus Corporation of the Americas · initiated Aug 6, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for undetected, deformed a-traumatic tips.

FDA root cause category: Process design

What the company is doing

Olympus notified consignees on about 08/06/2025 via letter. Consignees were instructed to examine inventory and quarantine any affected units, ensure all users of the device carefully read the content of this notification, including the reinforced text from the IFU and the product removal information, arrange for the return of affected units, and acknowledge receipt of the recall notification via the Olympus web portal. Consignees were also instructed to notify customers if the affected units were further distributed. ***Update 12/31/2025*** URGENT MEDICAL DEVICE RECALL letters were sent to additional customers on January 9, 2026 with identical instructions. ***Edit 3/23/26: January 9, 2026 letters users were instructed to return all affected devices.

Recall details

Recall number
Z-2484-2025
Status
Open, Classified
Initiated by the company
Aug 6, 2025
Posted by the FDA
Sep 5, 2025
Recalling firm
Olympus Corporation of the Americas
Quantity
104,508 units
Product code
KTI
FDA submissions
K193517
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Germany, Canada, Chile, China, Hong Kong, India, Japan, Singapore.

Product description

ViziShot 2FLEX. Single Use Aspiration Needle. Model Number: NA-U403SX-4019. The ViziShot 2 FLEX tissue biopsy device is made up of four main parts: a handle, a sheath, a needle, and a stylet. The sheath and needle are attached to the handle, and the stylet, which is removable, is located inside the needle tube.

Code information (models, lots, serial numbers)

Model Number: NA-U403SX-4019. UDI-DI: 00821925043060. Lot numbers: All lots manufactured after 7/29/2022.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates