Medline ReNewal Reprocessed St. Jude Medical Response and… recallZ-2460-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Jul 14, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Specific lots of reprocessed electrophysiology catheters may contain small residual particulates.

FDA root cause category: Under Investigation by firm

What the company is doing

An Urgent Medical Device Recall notification letter dated 7/14/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product immediately. 2. Please complete and return the enclosed response form via email to: [email protected] or fax: 866-767-1290, listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Upon receipt of your submitted response form, return labels and RGA information will be provided, if applicable. Your account will receive credit when the returned product is received. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request your customers return any affected product to you for collection and return to Medline Industries, LP. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2460-2025
Status
Open, Classified
Initiated by the company
Jul 14, 2025
Posted by the FDA
Aug 28, 2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
398 units
Product code
NLH
FDA submissions
K241156
Distribution
US Nationwide.

Product description

Medline ReNewal Reprocessed St. Jude Medical Response and Supreme Diagnostic Electrophysiology Catheter Models 401150RH, 401206RH, 401207RH, 401210RH, 401211RH, 401212RH, 401222RH, 401223RH, 401226RH, 401227RH, 401228RH, 401260RH, 401261RH, 401271RH, 401305RH, 401306RH, 401309RH, 401310RH, 401311RH, 401312RH, 401317RH, 401353RH, 401357RH, 401381RH, 401392RH, 401399RH, 401400RH, 401425RH, 401430RH, 401433RH, 401434RH, 401435RH, 401436RH, 401438RH, 401441RH, 401442RH, 401445RH, 401443RH, 401444RH, 401448RH, 401449RH, 401450RH, 401451RH, 401453RH, 401466RH, 401468RH, 401474RH, 401475RH, 401859RH, 401860RH, 401863RH, 401864RH, 401865RH, 401876RH, 401877RH, 401878RH, 401890RH, 401891RH, 401892RH, 401893RH, 401966RH, 401967RH, 401979RH, 401994RH, 401996RH, 402004RH, 402008RH, 402009RH, 402010RH, 402011RH, 402012RH

Code information (models, lots, serial numbers)

Show the full code information (5k characters)

1) 401211RH, UDI-DI: 10197344043736(each), 10197344043736(case), Lot Number: EP250106; 2) 401222RH, UDI-DI: 10197344043712(each), 10197344043712(case), Lot Number: EP241118; 3) 401222RH, UDI-DI: 10197344043712(each), 10197344043712(case), Lot Number: EP250106; 4) 401223RH, UDI-DI: 10197344043705(each), 10197344043705(case), Lot Number: EP250106; 5) 401226RH, UDI-DI: 10197344043699(each), 10197344043699(case), Lot Number: EP241118; 6) 401226RH, UDI-DI: 10197344043699(each), 10197344043699(case), Lot Number: EP250106; 7) 401227RH, UDI-DI: 10197344043682(each), 10197344043682(case), Lot Number: EP250106; 8) 401260RH, UDI-DI: 10197344043668(each), 10197344043668(case), Lot Number: EP241118; 9) 401260RH, UDI-DI: 10197344043668(each), 10197344043668(case), Lot Number: EP250106; 10) 401261RH, UDI-DI: 10197344043651(each), 10197344043651(case), Lot Number: EP250106; 11) 401305RH, UDI-DI: 10197344043637(each), 10197344043637(case), Lot Number: EP250106; 12) 401309RH, UDI-DI: 10197344043613(each), 10197344043613(case), Lot Number: EP250106; 13) 401353RH, UDI-DI: 10197344043569(each), 10197344043569(case), Lot Number: EP241118; 14) 401353RH, UDI-DI: 10197344043569(each), 10197344043569(case), Lot Number: EP250106; 15) 401381RH, UDI-DI: 10197344043545(each), 10197344043545(case), Lot Number: EP241118; 16) 401381RH, UDI-DI: 10197344043545(each), 10197344043545(case), Lot Number: EP250106; 17) 401399RH, UDI-DI: 10197344043521(each), 10197344043521(case), Lot Number: EP241118; 18) 401399RH, UDI-DI: 10197344043521(each), 10197344043521(case), Lot Number: EP250106; 19) 401430RH, UDI-DI: 10197344043491(each), 10197344043491(case), Lot Number: EP241118; 20) 401430RH, UDI-DI: 10197344043491(each), 10197344043491(case), Lot Number: EP250106; 21) 401433RH, UDI-DI: 10197344043484(each), 10197344043484(case), Lot Number: EP241118; 22) 401434RH, UDI-DI: 10197344043477(each), 10197344043477(case), Lot Number: EP241118; 23) 401435RH, UDI-DI: 10197344043460(each), 10197344043460(case), Lot Number: EP241118; 24) 401436RH, UDI-DI: 10197344043453(each), 10197344043453(case), Lot Number: EP250106; 25) 401441RH, UDI-DI: 10197344043439(each), 10197344043439(case), Lot Number: EP241118; 26) 401441RH, UDI-DI: 10197344043439(each), 10197344043439(case), Lot Number: EP250106; 27) 401442RH, UDI-DI: 10197344044245(each), 10197344044245(case), Lot Number: EP241118; 28) 401442RH, UDI-DI: 10197344044245(each), 10197344044245(case), Lot Number: EP250106; 29) 401443RH, UDI-DI: 10197344044252(each), 10197344044252(case), Lot Number: EP241118; 30) 401443RH, UDI-DI: 10197344044252(each), 10197344044252(case), Lot Number: EP250106; 31) 401444RH, UDI-DI: 10197344044269(each), 10197344044269(case), Lot Number: EP241118; 32) 401444RH, UDI-DI: 10197344044269(each), 10197344044269(case), Lot Number: EP250106; 33) 401448RH, UDI-DI: 10197344044283(each), 10197344044283(case), Lot Number: EP250106; 34) 401449RH, UDI-DI: 10197344044290(each), 10197344044290(case), Lot Number: EP241118; 35) 401449RH, UDI-DI: 10197344044290(each), 10197344044290(case), Lot Number: EP250106; 36) 401450RH, UDI-DI: 10197344044306(each), 10197344044306(case), Lot Number: EP241118; 37) 401450RH, UDI-DI: 10197344044306(each), 10197344044306(case), Lot Number: EP250106; 38) 401451RH, UDI-DI: 10197344044313(each), 10197344044313(case), Lot Number: EP241118; 39) 401451RH, UDI-DI: 10197344044313(each), 10197344044313(case), Lot Number: EP250106; 40) 401453RH, UDI-DI: 10197344044320(each), 10197344044320(case), Lot Number: EP250106; 41) 401474RH, UDI-DI: 10197344044351(each), 10197344044351(case), Lot Number: EP241118; 42) 401474RH, UDI-DI: 10197344044351(each), 10197344044351(case), Lot Number: EP250106; 43) 401860RH, UDI-DI: 10197344044382(each), 10197344044382(case), Lot Number: EP241118; 44) 401860RH, UDI-DI: 10197344044382(each), 10197344044382(case), Lot Number: EP250106; 45) 401863RH, UDI-DI: 10197344044399(each), 10197344044399(case), Lot Number: EP241118; 46) 401863RH, UDI-DI: 10197344044399(each), 10197344044399(case), Lot Number: EP250106; 47) 401864RH, UDI-DI: 10197344044405(each), 10197344044405(case), Lot Number: EP241118; 48) 401864RH, UDI-DI: 10197344044405(each), 10197344044405(case), Lot Number: EP250106; 49) 401877RH, UDI-DI: 10197344044436(each), 10197344044436(case), Lot Number: EP241118; 50) 401878RH, UDI-DI: 10197344044443(each), 10197344044443(case), Lot Number: EP241118; 51) 401890RH, UDI-DI: 10197344044450(each), 10197344044450(case), Lot Number: EP241118; 52) 401890RH, UDI-DI: 10197344044450(each), 10197344044450(case), Lot Number: EP250106; 53) 401891RH, UDI-DI: 10197344044467(each), 10197344044467(case), Lot Number: EP241118; 54) 401966RH, UDI-DI: 10197344044498(each), 10197344044498(case), Lot Number: EP250106; 55) 402004RH, UDI-DI: 10197344044566(each), 10197344044566(case), Lot Number: EP241118; 56) 402004RH, UDI-DI: 10197344044566(each), 10197344044566(case), Lot Number: EP250106; 57) 402008RH, UDI-DI: 10197344044573(each), 10197344044573(case), Lot Number: EP241118; 58) 402010RH, UDI-DI: 10197344044597(each), 10197344044597(case), Lot Number: EP241118; 59) 402010RH, UDI-DI: 10197344044597(each), 10197344044597(case), Lot Number: EP250106; 60) 402012RH, UDI-DI: 10197344043835(each), 10197344043835(case), Lot Number: EP241118; 61) 402012RH, UDI-DI: 10197344043835(each), 10197344043835(case), Lot Number: EP250106

Devices named in this recall

39 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K241156 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page