WATCHMAN recallZ-2436-2025 · Boston Scientific Corporation · initiated Jul 29, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
FDA root cause category: Under Investigation by firm
What the company is doing
Boston Scientific issued an Important Medical Device Advisory notice to its consignees on 07/29/2025 via FedEx. The notice explained the problem with the device, procedures under which the likelihood of the problem occurs, and risk to the patient. The WATCHMAN Access Systems Instructions for Use (IFUs) and WATCHMAN physician training will be updated to emphasize instructions related to Access System air management. This update will strengthen the information provided to clinicians regarding the potential for air embolism during WATCHMAN procedures performed under conscious or deep sedation and provide potential mitigation strategies. No devices are being removed. Instructions: 1. Review IFU updates related to air embolism as detailed in Appendix 2. These updates will be added to the WATCHMAN Access Systems IFUs and WATCHMAN Physician Training. 2. Forward this letter to any other clinicians in your medical facility who perform WATCHMAN procedures and to any facilities where affected devices have been transferred, including hospitals or sites within your network. If you are a distributor, this notice must be forwarded to your customers to ensure notification of this Medical Device Advisory is carried out to the end-user level. 3. Complete and return the enclosed Acknowledgment Form. If additional assistance is required or more information regarding this communication, customer should contact your local Boston Scientific representative or Sr. Quality Systems Manager at 763-494-1133.
Recall details
- Recall number
- Z-2436-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 29, 2025
- Posted by the FDA
- Aug 29, 2025
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 340185 units in total
- Product code
- DQY
- FDA submissions
- K240018
- Distribution
- Worldwide.
Product description
WATCHMAN TruSteer Access System, US, Material Number (UPN) M635TU90050;intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
Code information (models, lots, serial numbers)
GTIN 00191506022310, ALL NON-EXPIRED BATCHES
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97363).
- Z-2431-2025 · Open, ClassifiedWATCHMAN TruSeal Access System SGL, US, Material Number (UPN) M635TU70010; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Device…
- Z-2432-2025 · Open, ClassifiedWATCHMAN TruSeal Access System DBL, US, Material Number (UPN) M635TU70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Device…
- Z-2433-2025 · Open, ClassifiedWATCHMAN TruSeal Access System ANT, US, Material Number (UPN) M635TU70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Device…
- Z-2434-2025 · Open, ClassifiedWATCHMAN FXD Curve Access System Sgl, US, Material Number (UPN) M635TU80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with De…
- Z-2435-2025 · Open, ClassifiedWATCHMAN FXD Curve Access System Dbl, US, Material Number (UPN) M635TU80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with De…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page