TOPS Inserter, part of the TOPS System Instrument Set used… recallZ-2516-2025 · Premia Spine Ltd · initiated Jul 16, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Premia Spine Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for missing pins at tip of inserter.

FDA root cause category: Process change control

What the company is doing

On 07/16/2025, Premia Spine Ltd. notified Premia Spine Inc. via phone call. It was requested to quarantine any suspected reworked instruments that are at the warehouse and to initiate a field action and return the affected units of the TOPS Inserter that underwent rework. No customers (hospitals/physicians) were asked to quarantine or return units as all affected units were within Premia Spine's control, however one surgeon was notified, on about 07/31/2025, of the issue and was requested to review the surgical procedure, analyze relevant patient data, including x-ray, and perform additional actions if needed to confirm the pin has not remained in the patient's body.

Recall details

Recall number
Z-2516-2025
Status
Open, Classified
Initiated by the company
Jul 16, 2025
Posted by the FDA
Sep 4, 2025
Recalling firm
Premia Spine Ltd
Quantity
30 units
Product code
QWK
FDA submissions
P220002
Distribution
US Nationwide distribution in the state of Connecticut.

Product description

TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.

Code information (models, lots, serial numbers)

Model Number: 82889. UDI-DI: 07290115932638. Lot Numbers: RS20220546-1, RS20222162-1, RS20171777-1, RS20171884-1, RS0431-13B-1, 20152764-1, RS0431-02-1.