RayStation with the following product descriptions recallZ-2494-2025 · Raysearch Laboratories AB · initiated Aug 8, 2025
Class II Open, Classified All recalls posted in September 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Raysearch Laboratories AB or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
FDA root cause category: Under Investigation by firm
What the company is doing
RaySearch notified consignees on about 08/08/2025 via email. Consignees were informed that as temporary solution, affected objects can be re-exported via RayGateway to generate new UIDs. They were instructed to inform planning staff and all users about this workaround, inspect your product and identify all installed units with the above software version number(s), and complete and return the response form. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
Recall details
- Recall number
- Z-2494-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 8, 2025
- Posted by the FDA
- Sep 2, 2025
- Recalling firm
- Raysearch Laboratories AB
- Quantity
- 0
- Product code
- MUJ
- FDA submissions
- K171536
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
Product description
RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1, Software Version: 6.1.1.2. 2. RayStation 6 Service Pack 2, Software Version: 6.2.0.7. 3. RayStation 6 Service Pack 3, Software Version: 6.3.0.6. Product Description: Radiation Therapy Treatment Planning System.
Code information (models, lots, serial numbers)
1. Software Version: 6.1.1.2, UDI: 0735000201008220170529. 2. Software Version: 6.2.0.7, UDI: 0735000201007520170630. 3. Software Version: 6.3.0.6, UDI: 0735000201024220190923. Expiration Date: 2024-10-18
Other devices under the same FDA submission
2 devices cleared under K171536 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- RayStation 6.2Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment7 recalls0 reports
- RayStation 6.1Raysearch Laboratories AB (Publ) · System,Planning,Radiation Therapy Treatment2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97400).
- Z-2487-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 7, Software Version: 7.0.0.19. 2. RayStation 8A, Software Version: 8.0.0.61. 3. RayStation 8A Service Pack 1, …
- Z-2488-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 8B, Software Version: 8.1.0.47. 2. RayStation 8B Service Pack 1, Software Version: 8.1.1.8. 3. RayStation 8B S…
- Z-2489-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2. RayStation 9B Service Pack 1, Software Version: 9.2.0.483. 3. RayStation 1…
- Z-2491-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStatio…
- Z-2492-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 11B, Software Version: 12.0.0.932. 2. RayStation 11B Service Pack 1, Software Version: 12.1.0.1221. 3. RayStat…
- Z-2493-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 12A, Software Version: 13.0.0.1547. 2. RayStation 12A Service Pack 1, Software Version: 13.1.0.144. 3. RayStat…
- Z-2495-2025 · Open, ClassifiedRayStation with the following product descriptions: 1. RayStation 2023B, Software Version: 14.0.0.3338. 2. RayStation 2024A, Software Version: 15.0.0.430. 3. RayStation 2024A S…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page